Cobalt XT HF CDT-D MRI SureScan, Model Number: DTPA2D4, Implantable Cardioverter Defibrillator

Published July 3, 2024 · Recall initiated: April 16, 2024

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Potential for manufacturing defect. Weld crack led to devices failing at the leak check step in manufacturing.

What to do

Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit

Recall details

Manufacturer
Medtronic Inc.
Distributor
Worldwide - US Nationwide distribution in the states of CA, CT, FL, IN, MA, MI, NC, NH, NJ, NM, NY, OH, TX, VA and the countries of Australia, Austria, Bulgaria, Canada, China, Croatia, Czech Republic, France, Georgia, Germany, Greece, Guadeloupe, Iceland, Ireland, Italy, Jordan, Kosovo, Latvia, Moldova, Republic Of, Netherlands, Northern Ireland, Norway, Oman, Panama, Poland, Portugal, Qatar, Reunion, Slovakia, Slovenia, Spain, Sweden, Switzerland, United Arab Emirates, United Kingdom.
Lot
GTIN: 00763000178130, Serial Numbers: RTG612496S, RTG612497S, RTG612498S, RTG612499S, RTG612502S, RTG612503S, RTG612504S, RTG612505S, RTG612509S, RTG612512S, RTG612513S, RTG612549S, RTG612553S, RTG612556S.
Risk classification
Class II