Synthes(R) Hemostatic Bone Putty, Part numbers 08.901.001.97S, 08.901.001.98S, 08.901.001.99S, 08.901.001D, and VB1025.10S. Indicated for use as a water soluble implant material and for use in the control of bleeding from bone surfaces.
Published August 22, 2012 · Recall initiated: July 5, 2012
Summary
FDA records medical device Class I recall Z-2206-2012, published August 22, 2012 and initiated July 5, 2012. Recalling firm: Synthes USA HQ, Inc.; Status: Terminated; Reported quantity: 8,853 packs; Lot: All lot numbers associated with the identified part numbers. Class I means a reasonable probability of serious adverse health consequences or death.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
There is the potential for Hemostatic Bone Putty to ignite if contacted with electrosurgical cautery systems under certain conditions during surgery.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Synthes USA HQ, Inc.
- Distributor
- Worldwide Distribution -- US, including Washington, DC and states of AK, AZ, CA, CT, FL, GA, ID, KS, LA, MD, MA, MI, MN, MO, NE, NH, NY, NC, OH, OR, PA, SC, TX, VA, WA, and WI, and the country of Canada.
- Lot
- All lot numbers associated with the identified part numbers.
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDASynthes Hand Switch - Electric Pen Drive, Part Number 05.001.012 The Synthes Electric Pen Drive is indicatedJanuary 8, 2014
- FDASynthes TomoFix Plates The Synthes J5606-C TomoFix Osteotomy System is used for open and closed wedgeDecember 18, 2013
- FDASynthes Trochanteric Fixation Nail As part of the Synthes Trochanteric Fixation Nail (TFN) System, the TFNDecember 18, 2013
- FDASynthes Soft Tissue Retractor Small Extendible. Offset blade facilitates easy preparation of the epipereostealDecember 11, 2013
- FDAGE Healthcare Advantage Workstations, Advantage Worksation version 4.2 and 4.3 with Volume Viewer 2 (versionAugust 22, 2012
- FDAPhilips Essenta DR, X-Ray System, Diagnostic, general-purpose Model: 712070. The Essenta DR is aAugust 22, 2012
Questions about this recall
Was Synthes(R) Hemostatic Bone Putty, Part numbers 08.901.001.97S, 08.901.001.98S, 08.901.001.99S, 08.901.001D, and VB1025.10S. Indicated for use as a water soluble implant material and for use in the control of bleeding from bone surfaces. recalled?
Yes. FDA published recall Z-2206-2012 on August 22, 2012. Reported hazard: There is the potential for Hemostatic Bone Putty to ignite if contacted with electrosurgical cautery systems under certain conditions during surgery.
What should I do if I have this product?
Voluntary: Firm initiated · Letter