PrisMARS Kit (component of MARS Treatment Kit Type 1115/1 PrisMARS US), 1 MARS Treatment Kit per carton box, Product Number 800480, Batch
Published August 20, 2014 · Recall initiated: July 18, 2014
Summary
FDA records medical device Class II recall Z-2206-2014, published August 20, 2014 and initiated July 18, 2014. Recalling firm: Gambro Renal Products, Incorporated; Status: Terminated; Reported quantity: 401,451; Lot: Model Number MARS Treatment Kit Type 1115/1 PrisMARS US, Batch 0000017412 - 0000019827, Expiry Date 06/2014 to 05/2016. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Gambro initiated a field action on the Prisma and Prismaflex sets due to breakages of connectors resulting in leakage of liquid or blood.
What to do
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Gambro Renal Products, Incorporated
- Distributor
- Worldwide Distribution.
- Lot
- Model Number MARS Treatment Kit Type 1115/1 PrisMARS US, Batch 0000017412 - 0000019827, Expiry Date 06/2014 to 05/2016
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAPrisma M 100 PRE set. Product number 103657. Each set is packaged in a plastic pouch, 4 pouches per carton boxAugust 20, 2014
- FDAPrisma M 60 PRE set (new design). Product number 103658. Each set is packaged in a plastic pouch, 4 pouchesAugust 20, 2014
- FDAPrisma TPE 2000 set. Product number 109672. Each set is packaged in a plastic pouch, 4 pouches per carton boxAugust 20, 2014
- FDAPrismaflex M150. Product number 109990. Each set is packaged in a plastic pouch, 4 pouches per carton boxAugust 20, 2014
- FDAPrismaflex HF 1400. Product number 107142. Each set is packaged in a plastic pouch, 4 pouches per carton boxAugust 20, 2014
- FDAX-MARS Set (component of MARS Treatment Kit Type 1116/1 X-MARS US), Product Number 800541August 20, 2014
Questions about this recall
Was PrisMARS Kit (component of MARS Treatment Kit Type 1115/1 PrisMARS US), 1 MARS Treatment Kit per carton box, Product Number 800480, Batch recalled?
Yes. FDA published recall Z-2206-2014 on August 20, 2014. Reported hazard: Gambro initiated a field action on the Prisma and Prismaflex sets due to breakages of connectors resulting in leakage of liquid or blood.
What should I do if I have this product?
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit