VIDAS D-Dimer Exclusion II, an automated quantitative test for use on instruments of the VIDAS family for the immunoenzymatic determination of fibrin degradation products in human plasma using the Enzyme Linked Fluorescent Assay technique. Also indicated for use in conjunction with clinical pretest probability assessment model to exclude DVT and PE disease.
Published August 20, 2014 · Recall initiated: April 17, 2013
Summary
FDA records medical device Class II recall Z-2232-2014, published August 20, 2014 and initiated April 17, 2013. Recalling firm: BioMerieux SA; Status: Terminated; Reported quantity: 2160 units; Lot: Reference 304551-01 - Lots 1001625210, Exp. date 8/8/20123; 1001715580, Exp. date 9/26/2013; 1001711530, Exp. date 9/11/2013; 1001789540, Exp. date 10/29/2013; 1001932680, Exp. date 12/12/2013. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Customer complaints about not repeatable results below the detection limit (<45 ng/mL). Therefore false negatives are being reported for results around the medical decision point of 500 ng/mL that initially tested at <45 ng/mL.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- BioMerieux SA
- Distributor
- AL, AR, CA, CO, CT, DE, FL, GA, ID, IL, IN, IA, KS, KY, LA, ME, MA, MI, MN, MO, MT, NE, NH, NJ, NY, NC, OH, OK, OR, PA, RI, SC, TN, TX, VT, VA, WA, WV, WI, WY
- Lot
- Reference 304551-01 - Lots 1001625210, Exp. date 8/8/20123; 1001715580, Exp. date 9/26/2013; 1001711530, Exp. date 9/11/2013; 1001789540, Exp. date 10/29/2013; 1001932680, Exp. date 12/12/2013
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- France
Related recalls
- FDAEMAG system Model/Catalog Number: 418591/418591U Product Description: Nucleic acid sample preparationNovember 4, 2026
- FDAMYLA software. Used to manage microbiology test workflow from the reception of requests to the transmission ofAugust 18, 2021
- FDAVIDAS FSH, Ref 30407-01, IVD, Rx. The firm name on the label is bioMerieux SA, France. For use on theJuly 4, 2018
- FDAThe VITEK® 2 Gram Positive Susceptibility Card is intended for use with the VITEK 2 Systems in clinicalJune 20, 2018
- FDA23.5 Deg Rasp Handle Assy- RIGHT NonsterileAugust 20, 2014
- FDABD Multi Check CD4 Low Control Catalog No. 340916 is intended as a complete process control forAugust 20, 2014
Questions about this recall
Was VIDAS D-Dimer Exclusion II, an automated quantitative test for use on instruments of the VIDAS family for the immunoenzymatic determination of fibrin degradation products in human plasma using the Enzyme Linked Fluorescent Assay technique. Also indicated for use in conjunction with clinical pretest probability assessment model to exclude DVT and PE disease. recalled?
Yes. FDA published recall Z-2232-2014 on August 20, 2014. Reported hazard: Customer complaints about not repeatable results below the detection limit (<45 ng/mL). Therefore false negatives are being reported for results around the medical decision point of 500 ng/mL that initially tested at <45 ng/mL.
What should I do if I have this product?
Voluntary: Firm initiated · Letter