FDAFDA · Food and drugs·Recall number Z-2240-2024
ARTIS is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.
Published July 10, 2024 · Recall initiated: June 26, 2024
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
A potential issue with ARTIS icono systems equipped with a small detector and a corresponding collimator (model # 10843101) was identified. In very rare situations, the first x-ray release following patient registration may be performed with incorrect copper prefiltration. If the described issue occurs, the applied radiation dose rate may be higher than intended by the user.
What to do
FDA Mandated · Letter
Recall details
- Manufacturer
- Siemens AG/Siemens Healthcare GmbH
- Distributor
- Worldwide - US Nationwide distribution in the states of TX, MA, NH, MN, OK, NJ, SC, NY, MI, IL, MO, NC.
- Lot
- ARTIS icono biplane 11327600 4056869063317 ARTIS icono ceiling 11328100 4056869295923 ARTIS icono floor 11327700 4056869149325
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- Germany