ARTIS is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.
Published July 10, 2024 · Recall initiated: June 26, 2024
Summary
FDA records medical device Class II recall Z-2240-2024, published July 10, 2024 and initiated June 26, 2024. Recalling firm: Siemens AG/Siemens Healthcare GmbH; Status: Ongoing; Reported quantity: 201 units in total; Lot: ARTIS icono biplane 11327600 4056869063317 ARTIS icono ceiling 11328100 4056869295923 ARTIS icono floor 11327700 4056869149325. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
A potential issue with ARTIS icono systems equipped with a small detector and a corresponding collimator (model # 10843101) was identified. In very rare situations, the first x-ray release following patient registration may be performed with incorrect copper prefiltration. If the described issue occurs, the applied radiation dose rate may be higher than intended by the user.
What to do
FDA Mandated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Siemens AG/Siemens Healthcare GmbH
- Distributor
- Worldwide - US Nationwide distribution in the states of TX, MA, NH, MN, OK, NJ, SC, NY, MI, IL, MO, NC.
- Lot
- ARTIS icono biplane 11327600 4056869063317 ARTIS icono ceiling 11328100 4056869295923 ARTIS icono floor 11327700 4056869149325
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- Germany
Related recalls
- FDAARTIS is a family of dedicated angiography systems developed for single and biplane diagnostic imaging andJuly 10, 2024
- FDAARTIS is a family of dedicated angiography systems developed for single and biplane diagnostic imaging andJuly 10, 2024
- FDAAvalon Ultrasound Transducer. Directs low-energy ultrasound beam toward the fetal heart and externallyJuly 10, 2024
- FDAAMT Initial Placement Dilator Set. Used to place gastrostomy devices.July 10, 2024
- FDABeckman Coulter DxI 9000 Access Immunoassay Analyzer is a discrete photometric chemistry analyzer for clinicalJuly 10, 2024
- FDAWOM Tube Set for lrrlgatlon, single-use- Intended to transport irrigation fluid to the patient in endoscopicJuly 10, 2024
Questions about this recall
Was ARTIS is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. recalled?
Yes. FDA published recall Z-2240-2024 on July 10, 2024. Reported hazard: A potential issue with ARTIS icono systems equipped with a small detector and a corresponding collimator (model # 10843101) was identified. In very rare situations, the first x-ray release following patient registration may be performed with incorrect copper prefiltration. If the described issue occurs, the applied radiation dose rate may be higher than intended by the user.
What should I do if I have this product?
FDA Mandated · Letter