Vanguard 360 Femoral Boss Reamer, Label: REF 32-488474 VANGUARD 360 1NSTRUMENT FEMORAL BOSS REAMER NON-STERILE, BIOMET ORTHOPEDICS 56 EAST BELL DRIVE P.O. BOX 587 WARSAW, IN 46581 USA

Published August 27, 2014 · Recall initiated: July 1, 2014

Summary

FDA records medical device Class II recall Z-2242-2014, published August 27, 2014 and initiated July 1, 2014. Recalling firm: Biomet, Inc.; Status: Terminated; Reported quantity: 461; Lot: Catalog Number: 32-48847 4 Lot Numbers: 022120, 022180, 162530, 232460, 475500 and 475510. 699240, 793350, 793570 Additional Scope: 065030, 095780, 164290, 235870, 253670, 273630, 308620, 418970, 420030, 421210, 570460, 585260, 619690, 708240, 711430, 726490, 816160, 896810, 929690, 953540, 968950. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The Vanguard 360 Femoral Boss Reamers were manufactured oversized to the print. The instrument may not fit into the mating component.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Biomet, Inc.
Distributor
California, IA, FL, MD, UT, SC, AL, IL, CO, MA, KS, GA, WA, NV, AR, NH, OK,Pennsylvania, New York, Texas, Virginia, New Jersey, North Carolina, Kentucky, Michigan, Ohio, Wisconsin, Louisiana, Arizona, Indiana, South Dakota, Missouri & Montana. In addition to Korea, B.V., Japan, Puerto Rico, UK, Chile, New Zealand, Australia, & Canada
Lot
Catalog Number: 32-48847 4 Lot Numbers: 022120, 022180, 162530, 232460, 475500 and 475510. 699240, 793350, 793570 Additional Scope: 065030, 095780, 164290, 235870, 253670, 273630, 308620, 418970, 420030, 421210, 570460, 585260, 619690, 708240, 711430, 726490, 816160, 896810, 929690, 953540, 968950
Risk classification
Class II
Area (FDA)
device
Country
United States