Back to search
FDAFDA · Food and drugs·Recall number Z-2243-2023

Ventralight ST w Echo PS 4.5" Circle A low profile, bioresorbable, coated, permanent mesh, with a pre-attached removable positioning system, designed for the reconstruction of soft tissue deficiencies during laparoscopic ventral hernia repair. Catalog Number: 5955450G (EU only)

Published August 2, 2023 · Recall initiated: April 25, 2023

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Balloon may not inflate as intended. If the balloon does not inflate during laparoscopic ventral hernia repair this may result in a delay or prolongation of the procedure, inadequate mesh placement, erosion/migration, pain and mesh infection

What to do

Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit

Recall details

Manufacturer
Davol, Inc.
Distributor
Worldwide distribution - US Nationwide and the countries of Belgium, Czech Republic, France, Germany, Ireland, Italy, Netherlands, Portugal, Slovakia, South Africa, Spain, Sweden, Switzerland, UK, India, Hong Kong, Greece, Chile, and Colombia.
Lot
UDI-DI: (01)00801741201950(17)240528(10) Lot Number HUGT1609
Risk classification
Class II
Area (FDA)
device
Country
United States