Berchtold Chromophare F300 Exam Light CHROMOPHARE examination lights are medical lights used for local illumination of the patients' body so that illnesses, injuries and disabilities can be diagnosed and treated
Published June 14, 2017 · Recall initiated: April 21, 2017
Summary
FDA records medical device Class II recall Z-2248-2017, published June 14, 2017 and initiated April 21, 2017. Recalling firm: Stryker Communications; Status: Terminated; Reported quantity: 57 units; Lot: Serial Numbers: 7735170-X68377, 7735170-X68397, 7735170-X14846, 7735170-X15688, 7735170-X14492, 7735170-X14495, 7735170-X14497, 7735170-X14501, 7735170-X14502, 7735170-X14504, 7735170-X14506, 7735170-X14507, 7735170-X14508, 7735170-X14510, 7735170-X68383, 7735170-X68384, 7735170-X68385, 7735170-X68386, 7735170-X68393, 7735170-X68394, 7735170-X68395, 7735170-X68396, 7735170-X68392, 7735170-X68400, 7735170-X68401, 7735170-X15607, 7735170-X15608, 7735170-X15609, 7735170-X15610, 7735170-X15611, 7735170-X15612, 7735170-X15613, 7735170-X15614, 7735170-X15615. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The exam light may have a tolerance issue with the adapter assembly, which could potentially not allow the snap ring to be seated correctly. If this is the case, there could be insufficient mount force that may cause the equipment to fall, resulting in serious injury.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Stryker Communications
- Distributor
- Worldwide Distribution - US and Canada
- Lot
- Serial Numbers: 7735170-X68377, 7735170-X68397, 7735170-X14846, 7735170-X15688, 7735170-X14492, 7735170-X14495, 7735170-X14497, 7735170-X14501, 7735170-X14502, 7735170-X14504, 7735170-X14506, 7735170-X14507, 7735170-X14508, 7735170-X14510, 7735170-X68383, 7735170-X68384, 7735170-X68385, 7735170-X68386, 7735170-X68393, 7735170-X68394, 7735170-X68395, 7735170-X68396, 7735170-X68392, 7735170-X68400, 7735170-X68401, 7735170-X15607, 7735170-X15608, 7735170-X15609, 7735170-X15610, 7735170-X15611, 7735170-X15612, 7735170-X15613, 7735170-X15614, 7735170-X15615
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAChromophare Surgical Light System, REF: CH00000001; SLX to Oculan NFC Upgrade Kit, REF: P60034October 1, 2025
- FDAChromophare Surgical Light System , REF CH00000001April 17, 2024
- FDABerchtold Chromophare Dual Monitor Carrier Arm or Dual Flat Panel, a component of Berchtold ChromophareSeptember 11, 2019
- FDAEDS Light Suspension, Central Axis, a component of the Stryker Visum LED Surgical Lighting System, with theMay 29, 2019
- FDAKYPHON(R) Directional Bone Filler Device, Size: 3, REF F04C, Rx Only, STERILE R.June 14, 2017
- FDAHudson RCI, Pediatric Nasal Cannula, Non-Flared Prongs, Rx only,June 14, 2017
Questions about this recall
Was Berchtold Chromophare F300 Exam Light CHROMOPHARE examination lights are medical lights used for local illumination of the patients' body so that illnesses, injuries and disabilities can be diagnosed and treated recalled?
Yes. FDA published recall Z-2248-2017 on June 14, 2017. Reported hazard: The exam light may have a tolerance issue with the adapter assembly, which could potentially not allow the snap ring to be seated correctly. If this is the case, there could be insufficient mount force that may cause the equipment to fall, resulting in serious injury.
What should I do if I have this product?
Voluntary: Firm initiated · Letter