Newport HT70 and HT70 Plus Ventilators, Model Number: HT70 and HT70 Plus. Product Usage: The HT70 ventilator system is intended to provide continuous or intermittent positive pressure mechanical ventilatory support for the care of individuals who require mechanical ventilation through invasive or non-invasive interfaces. Specifically, the HT70 family of ventilators is applicable for infant, pediatric and adult patients greater than or equal to 5 kg (11 lbs.). The HT70 is a restricted medical device intended for use by qualified, trained personnel under the direction of a physician. The HT70 is suitable for use in hospital, sub-acute, emergency room, and home care environments, as well as for transport and emergency response applications.
Published September 12, 2012 · Recall initiated: August 3, 2012
Summary
FDA records medical device Class I recall Z-2250-2012, published September 12, 2012 and initiated August 3, 2012. Recalling firm: Newport Medical Instruments Inc; Status: Terminated; Reported quantity: 72; Lot: Model Number: HT70 and HT70 Plus. Serial numbers: N12HT700514226 to N12HT700514231, N12HT700514233 to N12HT700514236, N12HT700514238 to N12HT700614280, N12HT700614282 to N12HT700614299, N12HT700614301 to N12HT700614309, N12HT720410266 to N12HT720410268, N12HT720410271, N12HT720410273, N12HT720410282, N12HT720410283, N12HT720410286, N12HT720410288, N12HT720410289, N12HT720410290, N12HT720410294 to N12HT720510300, N12HT720510302, N12HT720510322 to N12HT720510330, N12HT720510332 to N12HT720510344,. Class I means a reasonable probability of serious adverse health consequences or death.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
May emit a continuous high priority alarm and the ventilator may stop ventilating, due to a component failure on the control board.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Newport Medical Instruments Inc
- Distributor
- Worldwide distribution: USA (nationwide) and countries of: Brazil, Chile, Egypt, France, Hong Kong, India, Japan, Norway, Philippines and South Africa.
- Lot
- Model Number: HT70 and HT70 Plus. Serial numbers: N12HT700514226 to N12HT700514231, N12HT700514233 to N12HT700514236, N12HT700514238 to N12HT700614280, N12HT700614282 to N12HT700614299, N12HT700614301 to N12HT700614309, N12HT720410266 to N12HT720410268, N12HT720410271, N12HT720410273, N12HT720410282, N12HT720410283, N12HT720410286, N12HT720410288, N12HT720410289, N12HT720410290, N12HT720410294 to N12HT720510300, N12HT720510302, N12HT720510322 to N12HT720510330, N12HT720510332 to N12HT720510344,
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDANewport Medical Instruments HT70 and HT70 Plus ventilator Power Pac batteries Model number: HT70 and HT70May 8, 2013
- FDAHumphrey Chocolate Drizzle, 5 oz. in a clear plastic bag, UPC Code 8108400009. Bags are packed 12 to a case.September 12, 2012
- FDAV60 Ventilator, Continuous Ventilator, Model# V60, PN 1056490 Product Usage: The V60 Ventilator is an assistSeptember 12, 2012
- FDAHumphrey Candy Kisses, 4 oz. individually wrapped and contained in a clear plastic bag, UPC Code 81084 00002.September 12, 2012
- FDAsyngo(R) Lab Data Manager and syngo(R) Lab Process Manager, all versions. This device is classified as a dataSeptember 12, 2012
- FDAFresh Express Hearts of Romaine Crunchy and Sweet 100% Hearts of Romaine Lettuce. 10 oz. plastic bag UPC 71279September 12, 2012
Questions about this recall
Was Newport HT70 and HT70 Plus Ventilators, Model Number: HT70 and HT70 Plus. Product Usage: The HT70 ventilator system is intended to provide continuous or intermittent positive pressure mechanical ventilatory support for the care of individuals who require mechanical ventilation through invasive or non-invasive interfaces. Specifically, the HT70 family of ventilators is applicable for infant, pediatric and adult patients greater than or equal to 5 kg (11 lbs.). The HT70 is a restricted medical device intended for use by qualified, trained personnel under the direction of a physician. The HT70 is suitable for use in hospital, sub-acute, emergency room, and home care environments, as well as for transport and emergency response applications. recalled?
Yes. FDA published recall Z-2250-2012 on September 12, 2012. Reported hazard: May emit a continuous high priority alarm and the ventilator may stop ventilating, due to a component failure on the control board.
What should I do if I have this product?
Voluntary: Firm initiated · Letter