ON-Q pump with ONDEMAND bolus button, ON-Q C-bloc with ONDEMAND; ON-Q C-bloc with Select a Flow and ONDEMAND Model/Reference #: 7000180, 7000181, 7000182, 7000367, 7000397, CB003, CB005, CB006, P400X1-7+5/60, P400X2-14+5/30, P400X5+5/60, and PMB01. Part Numbers: 101347100, 101347300, 101347400, 101360200, 101368400, 101368600, 101368700, 102987000, 102987200, 102987300, 103161800, 103534300, 5001438, 5001470, 5001651, 5001780, 5001808, 5001809, 5001812, 7000180, 7000181, 7000182, & 7000367. Indicated for continuous and or intermittent delivery of medication to or around surgical wound sites and or close proximity to nerves.
Published September 5, 2012 · Recall initiated: May 8, 2012
Summary
FDA records medical device Class I recall Z-2253-2012, published September 5, 2012 and initiated May 8, 2012. Recalling firm: I-Flow Corporation; Status: Terminated; Reported quantity: 80,646 units total (65,926 units in US); Lot: Products having a 10-digit Lot # that is greater than 0200521454 (i.e. 0200521455 and greater) are not impacted by this recall. Class I means a reasonable probability of serious adverse health consequences or death.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
It was determined that in a small quantity of ON-Q pumps with ONDEMAND, the bolus button may not lock in the down position when depressed and/or the orange bolus refill indicator may stay in the lowermost position. As a result, the patient may receive continuous infusion at a rate greater than expected.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- I-Flow Corporation
- Distributor
- Worldwide Distribution -- USA, Canada, Australia, the EU (Germany), United Arab Emirates, and Cyprus.
- Lot
- Products having a 10-digit Lot # that is greater than 0200521454 (i.e. 0200521455 and greater) are not impacted by this recall.
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAOEC IT3000; IT2500 (EnTrak); IT2500 Plus (EnTrak Plus); IT3500 (InstaTrak); IT3500 Plus (InstaTrak Plus);September 5, 2012
- FDACisco brand HealthPresence, version 2.1; Model/Catalog Number: CHP-ENTPRSYR2.I-K9; Product is manufactured andSeptember 5, 2012
- FDANaturally Fresh Salad Toppings, Roasted & Glazed Pecan Pieces, Net weight 3.5 oz, Distributed by:September 5, 2012
- FDAsyngo Imaging XS. Radiological image processing system.September 5, 2012
- FDAAussie-Style Lucky Country Soft Licorice in 1.5 lb. bags. The product is packaged 1.5lb x 18 bags per tray.September 5, 2012
- FDACanon Solid State X-ray Imager (Flat Panel/Digital Imager), Digital Radiography, Models CXDI-55C, CXDI-55G andSeptember 5, 2012
Questions about this recall
Was ON-Q pump with ONDEMAND bolus button, ON-Q C-bloc with ONDEMAND; ON-Q C-bloc with Select a Flow and ONDEMAND Model/Reference #: 7000180, 7000181, 7000182, 7000367, 7000397, CB003, CB005, CB006, P400X1-7+5/60, P400X2-14+5/30, P400X5+5/60, and PMB01. Part Numbers: 101347100, 101347300, 101347400, 101360200, 101368400, 101368600, 101368700, 102987000, 102987200, 102987300, 103161800, 103534300, 5001438, 5001470, 5001651, 5001780, 5001808, 5001809, 5001812, 7000180, 7000181, 7000182, & 7000367. Indicated for continuous and or intermittent delivery of medication to or around surgical wound sites and or close proximity to nerves. recalled?
Yes. FDA published recall Z-2253-2012 on September 5, 2012. Reported hazard: It was determined that in a small quantity of ON-Q pumps with ONDEMAND, the bolus button may not lock in the down position when depressed and/or the orange bolus refill indicator may stay in the lowermost position. As a result, the patient may receive continuous infusion at a rate greater than expected.
What should I do if I have this product?
Voluntary: Firm initiated · Letter