GE Healthcare Discovery IGS 730 Cardiovascular X-Ray Imaging System. The angiography X-ray systems are indicated for use in generating fluoroscopic and rotational images of human anatomy for cardiovascular, vascular and non-vascular, diagnostic and interventional procedures. Additionally, with the OR table, the angiographic X-ray systems are indicated for use in generating fluoroscopic and rotational images of human anatomy for image guided surgical procedures. The OR table is suitable for interventional and surgical procedures.
Published September 25, 2013 · Recall initiated: June 19, 2013
Summary
FDA records medical device Class II recall Z-2253-2013, published September 25, 2013 and initiated June 19, 2013. Recalling firm: GE Healthcare, LLC; Status: Terminated; Reported quantity: 4 (2 US, 2 OUS); Lot: Mfg Lot or Serial # 00000611882BU2 00000612803BU7 00000613979BU4 00000620447BU3. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
GE Healthcare has recently become aware of a potential safety issue with respect to Discovery IGS 730 Imaging Systems. A patient data management error may prevent the Discovery IGS 730 Imaging System from booting on start-up or re-booting during procedure. The boot issue may occur when a large amount of patient data remains in the browser and is not deleted. The inability to boot the system ma
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- GE Healthcare, LLC
- Distributor
- Worldwide distribution: US (nationwide) including states of : PA and TN; countries of: FRANCE and SINGAPORE.
- Lot
- Mfg Lot or Serial # 00000611882BU2 00000612803BU7 00000613979BU4 00000620447BU3
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDACentricity PACS Software Version 7.0 SP0.0.4.7August 3, 2022
- FDASIGNA Premier magnetic resonance scanner, model 5748519.July 27, 2022
- FDAGE Centricity Universal Viewer Zero Footprint. To view, communicate, process, and display Medical images andJuly 13, 2022
- FDABack up batteries in Avance CS2 and Avance CS2 Pro anesthesia devices manufactured on or after April 1, 2019July 6, 2022
- FDACentral Dressing Change Kit; AMS-8464CP- 6. Vygon Central Dressing Change Kit contains: 1 Pair Nitrile GlovesSeptember 25, 2013
- FDAPOWERWAND Safety Introducer with Extended Dwell Catheter; Catalog no. 92006. Used to gain access to theSeptember 25, 2013
Questions about this recall
Was GE Healthcare Discovery IGS 730 Cardiovascular X-Ray Imaging System. The angiography X-ray systems are indicated for use in generating fluoroscopic and rotational images of human anatomy for cardiovascular, vascular and non-vascular, diagnostic and interventional procedures. Additionally, with the OR table, the angiographic X-ray systems are indicated for use in generating fluoroscopic and rotational images of human anatomy for image guided surgical procedures. The OR table is suitable for interventional and surgical procedures. recalled?
Yes. FDA published recall Z-2253-2013 on September 25, 2013. Reported hazard: GE Healthcare has recently become aware of a potential safety issue with respect to Discovery IGS 730 Imaging Systems. A patient data management error may prevent the Discovery IGS 730 Imaging System from booting on start-up or re-booting during procedure. The boot issue may occur when a large amount of patient data remains in the browser and is not deleted. The inability to boot the system ma
What should I do if I have this product?
Voluntary: Firm initiated · Letter