FDAFDA · Food and drugs·Recall number Z-2254-2012

Hitachi Echelon MRI System/Oasis MRI, Software. These are software controlled MRI devices. Product Usage: The Echelon and Oasis MR systems are imaging devices, intended to provide the physician with physiological and clinical information that can be useful in diagnosis determination.

Published August 29, 2012 · Recall initiated: June 15, 2012

Summary

FDA records medical device Class II recall Z-2254-2012, published August 29, 2012 and initiated June 15, 2012. Recalling firm: Hitachi Medical Systems America Inc; Status: Terminated; Reported quantity: 165 units; Lot: Echelon, V001 through V061, inclusive. Oasis, M001 through M105, inclusive, M951. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Hitachi discovered a software error with a feature called Multi-Planar Reconstruction (MPR). MPR takes an original volume dataset reconstructed in any plane view and creates a new image set from another view (i.e. axial to coronal, coronal to sagittal, etc.). The error causes the first image in the dataset to have a right/left reversed orientation. The software error can occur randomly when execu

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Hitachi Medical Systems America Inc
Distributor
US Nationwide Distribution - including the states of:: AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NJ, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, WY and Mexico.
Lot
Echelon, V001 through V061, inclusive. Oasis, M001 through M105, inclusive, M951
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Hitachi Echelon MRI System/Oasis MRI, Software. These are software controlled MRI devices. Product Usage: The Echelon and Oasis MR systems are imaging devices, intended to provide the physician with physiological and clinical information that can be useful in diagnosis determination. recalled?

Yes. FDA published recall Z-2254-2012 on August 29, 2012. Reported hazard: Hitachi discovered a software error with a feature called Multi-Planar Reconstruction (MPR). MPR takes an original volume dataset reconstructed in any plane view and creates a new image set from another view (i.e. axial to coronal, coronal to sagittal, etc.). The error causes the first image in the dataset to have a right/left reversed orientation. The software error can occur randomly when execu

What should I do if I have this product?

Voluntary: Firm initiated · Letter