Hitachi Echelon MRI System/Oasis MRI, Software. These are software controlled MRI devices. Product Usage: The Echelon and Oasis MR systems are imaging devices, intended to provide the physician with physiological and clinical information that can be useful in diagnosis determination.
Published August 29, 2012 · Recall initiated: June 15, 2012
Summary
FDA records medical device Class II recall Z-2254-2012, published August 29, 2012 and initiated June 15, 2012. Recalling firm: Hitachi Medical Systems America Inc; Status: Terminated; Reported quantity: 165 units; Lot: Echelon, V001 through V061, inclusive. Oasis, M001 through M105, inclusive, M951. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Hitachi discovered a software error with a feature called Multi-Planar Reconstruction (MPR). MPR takes an original volume dataset reconstructed in any plane view and creates a new image set from another view (i.e. axial to coronal, coronal to sagittal, etc.). The error causes the first image in the dataset to have a right/left reversed orientation. The software error can occur randomly when execu
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Hitachi Medical Systems America Inc
- Distributor
- US Nationwide Distribution - including the states of:: AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NJ, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, WY and Mexico.
- Lot
- Echelon, V001 through V061, inclusive. Oasis, M001 through M105, inclusive, M951
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAHitachi Echelon MRI SystemMarch 29, 2017
- FDAHitachi Oasis MRI SystemMarch 29, 2017
- FDAHitachi Oasis MRI System - C-Spine CoilFebruary 1, 2017
- FDAHitachi Echelon Oval MRI system is a diagnostic imaging device (one unit per package) and is intended toSeptember 10, 2014
- FDAAGFA IMPAC CV DICOMStore with Media Purge Daemon (MPD) and IMPAX CV DICOMStore with Cardiovascular PurgeAugust 29, 2012
- FDAUniCel DxI 800 Access Immunoassay Systems, Part Numbers: Dxl 800: 937100; Dxl 800 Remanufactured: A25288; DxlAugust 29, 2012
Questions about this recall
Was Hitachi Echelon MRI System/Oasis MRI, Software. These are software controlled MRI devices. Product Usage: The Echelon and Oasis MR systems are imaging devices, intended to provide the physician with physiological and clinical information that can be useful in diagnosis determination. recalled?
Yes. FDA published recall Z-2254-2012 on August 29, 2012. Reported hazard: Hitachi discovered a software error with a feature called Multi-Planar Reconstruction (MPR). MPR takes an original volume dataset reconstructed in any plane view and creates a new image set from another view (i.e. axial to coronal, coronal to sagittal, etc.). The error causes the first image in the dataset to have a right/left reversed orientation. The software error can occur randomly when execu
What should I do if I have this product?
Voluntary: Firm initiated · Letter