PS Tibial Insert, REF 163-1XXX Product Usage: UHMWPE component used in TKA procedures to serve as a tibial bearing surface for articulation with the femoral component
Published August 27, 2014 · Recall initiated: July 11, 2014
Summary
FDA records medical device Class II recall Z-2255-2014, published August 27, 2014 and initiated July 11, 2014. Recalling firm: Ortho Development Corporation; Status: Terminated; Reported quantity: 11; Lot: PS Tibial Insert Size 3 8mm,Part 163-1308,Lot A135943,PS Tibial Insert Size 3 9mm,Part 163-1309,Lot A136492,PS Tibial Insert Size 3 12mm,Part 163-1312,Lot A136453,PS Tibial Insert Size 4 7mm,Part 163-1407,Lot A135945,PS Tibial Insert Size 4 7mm,Part 163-1407,Lot A136454,PS Tibial Insert Size 5 9mm,Part 163-1509,Lot A135827,. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Ortho Development is recalling various Pivot Bipolar Cup, Patella, PS Tibial Insert, UC Tibila Insert, and CK Tibil Insert because the foil pouches did not properly seal, allowing air to seep back into the pouch after sealing which increases the risk that the material properties of the implant would be substantially degraded due to oxidation.
What to do
Voluntary: Firm initiated · E-Mail
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Ortho Development Corporation
- Distributor
- Worldwide Distribution - US Nationwide and the country of Japan. No military/govt/VA consignees.
- Lot
- PS Tibial Insert Size 3 8mm,Part 163-1308,Lot A135943,PS Tibial Insert Size 3 9mm,Part 163-1309,Lot A136492,PS Tibial Insert Size 3 12mm,Part 163-1312,Lot A136453,PS Tibial Insert Size 4 7mm,Part 163-1407,Lot A135945,PS Tibial Insert Size 4 7mm,Part 163-1407,Lot A136454,PS Tibial Insert Size 5 9mm,Part 163-1509,Lot A135827,
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDATibial Tray Nonporous A Size 2; Model # PN 162-1200-A; Material: Ti6Al4V.May 13, 2020
- FDATibila Insert UC, Balanced Knee System, REF163-3YYY Product Usage: UHMWPE component used in TKA procedures toAugust 27, 2014
- FDATibila Insert CK, Balanced Knee System Revision, REF563-1YYY Product Usage: UHMWPE component used in TKA andAugust 27, 2014
- FDAPivot Bipolar Cup 22x42/43-51/53mm, REF 133-YYYY Product Usage: Used in a hemiarthroplasty procedure as theAugust 27, 2014
- FDAPatella 38mm, REF164-0038 Product Usage: UHMWPE component used in a TKA procedure to resurface the patella andAugust 27, 2014
- FDAPhilips IntelliSpace PACS 4.4 image management systems performs digital image processing, measurement andAugust 27, 2014
Questions about this recall
Was PS Tibial Insert, REF 163-1XXX Product Usage: UHMWPE component used in TKA procedures to serve as a tibial bearing surface for articulation with the femoral component recalled?
Yes. FDA published recall Z-2255-2014 on August 27, 2014. Reported hazard: Ortho Development is recalling various Pivot Bipolar Cup, Patella, PS Tibial Insert, UC Tibila Insert, and CK Tibil Insert because the foil pouches did not properly seal, allowing air to seep back into the pouch after sealing which increases the risk that the material properties of the implant would be substantially degraded due to oxidation.
What should I do if I have this product?
Voluntary: Firm initiated · E-Mail