AUTOCOMP6 XPS High Speed Compounder REF 58810

Published August 3, 2016 · Recall initiated: June 23, 2016

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

An issue was identified after a redesign of the auto compounding device. The compounder cannot meet the designed accuracy specification when dispensing into bags. After reviewing the initial design of the device, it was determined the dispensing volume validation was not properly done and none of the compounders on the market can meet the stated specifications.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
The Metrix Company
Distributor
Product distributed to medical facilities in the following US states: AL, AZ, CA, CO, CT, FL, IA, IL, KS, LA, MA, MD, MN, NC, OH, PA, PR, TN, TX, UT, and WA. Product was also distributed internationally to the following countries: Australia, Canada, Dominican Republic, Malaysia, Turkey, and United Kingdom
Lot
000001, 003002, 003003, 003004, 003005, 003006, 003007, 003008, 003009, 003010, 003011, 003012, 003013, 003014, 003015, 003016, 003017, 003018, 003019, 003020, 003023, 003024, 003025, 003026, 003027, 003028, 003029, 003030, 003031, 003036, 003037, 003038, 003039, 003040, 003041, 003042, 003043, 003044, 003045, 003046, 003047, 003048, 003049, 003050, 003051, 003052, 003053, 003054, 003055, 003056, 003057
Risk classification
Class II
Area (FDA)
device
Country
United States