Neurostimulation Kit (Ported, Dual 8, 40 cm), REF: 71005-US Peripheral Nerve Stimulation Kit (Ported, Dual 8, 40 cm), REF: 71019-US Components of the Nalu Spinal Cord Stimulation System.

Published July 10, 2024 · Recall initiated: June 3, 2024

Summary

FDA records medical device Class II recall Z-2257-2024, published July 10, 2024 and initiated June 3, 2024. Recalling firm: Nalu Medical, Inc.; Status: Ongoing; Reported quantity: 13; Lot: UDI-DI: 71005-US: 00812537036137, 00812537033600, 00812537033518, 00812537030975, 00812537030524. 71019-US: 00812537036441, 00812537035604, 00812537035598 Serial Numbers: H01377, H01382, H01432, H01419, H01387, H01445, H01415, H01441, H01373, H01428, H01378, H01438, H01431. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Neurostimulation Systems have a manufacturing non-conformance that may reduce the efficiency of the radiofrequency signals between the Implantable Pulse Generator (IPG) and the Therapy Disc, causing the IPG to be non-functional, which causes the device to be non-functional so it will no longer inhibit the transmission of pain signals to the brain.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Nalu Medical, Inc.
Distributor
US Nationwide distribution in the states of MO, IL, CA, TX, TN, OK, AZ, AR.
Lot
UDI-DI: 71005-US: 00812537036137, 00812537033600, 00812537033518, 00812537030975, 00812537030524. 71019-US: 00812537036441, 00812537035604, 00812537035598 Serial Numbers: H01377, H01382, H01432, H01419, H01387, H01445, H01415, H01441, H01373, H01428, H01378, H01438, H01431
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Neurostimulation Kit (Ported, Dual 8, 40 cm), REF: 71005-US Peripheral Nerve Stimulation Kit (Ported, Dual 8, 40 cm), REF: 71019-US Components of the Nalu Spinal Cord Stimulation System. recalled?

Yes. FDA published recall Z-2257-2024 on July 10, 2024. Reported hazard: Neurostimulation Systems have a manufacturing non-conformance that may reduce the efficiency of the radiofrequency signals between the Implantable Pulse Generator (IPG) and the Therapy Disc, causing the IPG to be non-functional, which causes the device to be non-functional so it will no longer inhibit the transmission of pain signals to the brain.

What should I do if I have this product?

Voluntary: Firm initiated · Letter