da Vinci SP surgical system
Published June 10, 2020 · Recall initiated: April 1, 2019
Summary
FDA records medical device Class II recall Z-2260-2020, published June 10, 2020 and initiated April 1, 2019. Recalling firm: Intuitive Surgical, Inc.; Status: Terminated; Reported quantity: 17; Lot: Serial Numbers: SP0022, SP0031, SP0023, SP0032, SP0024, SP0034, SP0025, SP0035, SP0026, SP0036, SP0027, SP0037, SP0028, SP0038, SP0029, SP0039, SP0030. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Intuitive has become aware that the da Vinci SP system may trigger a mechanical vibration of the instrument tips and endoscope due to a software anomaly. This issue has the potential to occur only under the following, extremely rare specific conditions: 1) the user is activating Adjust Mode, AND 2) the instrument Arm is near its vertical position limit, AND 3) the user is applying sustained force against the hand control haptic feedback.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Intuitive Surgical, Inc.
- Distributor
- U.S.: OH, PN, NY, MN, TN, TX, OR, IL, MD, NJ, AL. O.U.S.: South Korea
- Lot
- Serial Numbers: SP0022, SP0031, SP0023, SP0032, SP0024, SP0034, SP0025, SP0035, SP0026, SP0036, SP0027, SP0037, SP0028, SP0038, SP0029, SP0039, SP0030
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAda Vinci SP Instrument Arm Drape, SP, 1-PACK, Part Number: 430013-13, 430013-15July 22, 2026
- FDADa Vinci ASSY, PSS, SP1098, Part Number: 380601July 1, 2026
- FDABrand Name: Da Vinci 5 Product Name: ASSY,DV5 TOWER,IS5000 Model/Catalog Number: 380746 Software Version: N/AJune 10, 2026
- FDAIntuitive 8 mm SureForm 30 Gray Reloads Reference Numbers: 48230M-05 and 48230M-06May 13, 2026
- FDAARTIS icono systems with Quantification Application SW (QVA/QCA) on a standalone XWP computer, anJune 10, 2020
- FDAPerkinElmer 226 Sample Collection DeviceJune 10, 2020
Questions about this recall
Was da Vinci SP surgical system recalled?
Yes. FDA published recall Z-2260-2020 on June 10, 2020. Reported hazard: Intuitive has become aware that the da Vinci SP system may trigger a mechanical vibration of the instrument tips and endoscope due to a software anomaly. This issue has the potential to occur only under the following, extremely rare specific conditions: 1) the user is activating Adjust Mode, AND 2) the instrument Arm is near its vertical position limit, AND 3) the user is applying sustained force against the hand control haptic feedback.
What should I do if I have this product?
Voluntary: Firm initiated · Letter