da Vinci SP surgical system

Published June 10, 2020 · Recall initiated: April 1, 2019

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Intuitive has become aware that the da Vinci SP system may trigger a mechanical vibration of the instrument tips and endoscope due to a software anomaly. This issue has the potential to occur only under the following, extremely rare specific conditions: 1) the user is activating Adjust Mode, AND 2) the instrument Arm is near its vertical position limit, AND 3) the user is applying sustained force against the hand control haptic feedback.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Intuitive Surgical, Inc.
Distributor
U.S.: OH, PN, NY, MN, TN, TX, OR, IL, MD, NJ, AL. O.U.S.: South Korea
Lot
Serial Numbers: SP0022, SP0031, SP0023, SP0032, SP0024, SP0034, SP0025, SP0035, SP0026, SP0036, SP0027, SP0037, SP0028, SP0038, SP0029, SP0039, SP0030
Risk classification
Class II