FDAFDA · Food and drugs·Recall number Z-2260-2020

da Vinci SP surgical system

Published June 10, 2020 · Recall initiated: April 1, 2019

Summary

FDA records medical device Class II recall Z-2260-2020, published June 10, 2020 and initiated April 1, 2019. Recalling firm: Intuitive Surgical, Inc.; Status: Terminated; Reported quantity: 17; Lot: Serial Numbers: SP0022, SP0031, SP0023, SP0032, SP0024, SP0034, SP0025, SP0035, SP0026, SP0036, SP0027, SP0037, SP0028, SP0038, SP0029, SP0039, SP0030. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Intuitive has become aware that the da Vinci SP system may trigger a mechanical vibration of the instrument tips and endoscope due to a software anomaly. This issue has the potential to occur only under the following, extremely rare specific conditions: 1) the user is activating Adjust Mode, AND 2) the instrument Arm is near its vertical position limit, AND 3) the user is applying sustained force against the hand control haptic feedback.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Intuitive Surgical, Inc.
Distributor
U.S.: OH, PN, NY, MN, TN, TX, OR, IL, MD, NJ, AL. O.U.S.: South Korea
Lot
Serial Numbers: SP0022, SP0031, SP0023, SP0032, SP0024, SP0034, SP0025, SP0035, SP0026, SP0036, SP0027, SP0037, SP0028, SP0038, SP0029, SP0039, SP0030
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was da Vinci SP surgical system recalled?

Yes. FDA published recall Z-2260-2020 on June 10, 2020. Reported hazard: Intuitive has become aware that the da Vinci SP system may trigger a mechanical vibration of the instrument tips and endoscope due to a software anomaly. This issue has the potential to occur only under the following, extremely rare specific conditions: 1) the user is activating Adjust Mode, AND 2) the instrument Arm is near its vertical position limit, AND 3) the user is applying sustained force against the hand control haptic feedback.