Container fill and transfer sets, automatic fill compatible, 250 mL, 500 mL, 1000 mL, 1500 mL, 2000 mL, 3000 mL, or 4000 mL.
Published August 3, 2016 · Recall initiated: June 23, 2016
Summary
FDA records medical device Class II recall Z-2261-2016, published August 3, 2016 and initiated June 23, 2016. Recalling firm: The Metrix Company; Status: Terminated; Reported quantity: 18,366 cases; Lot: All lots are being recalled. Lot code consists of the five digit product code followed by a dash followed by a unique five character numeric or alphanumeric sequence. Affected product codes for this product: 58740, 58741, 58742, 58749, 58750, 58753, 58754. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
An issue was identified after a redesign of the auto compounding device. The compounder cannot meet the designed accuracy specification when dispensing into bags. After reviewing the initial design of the device, it was determined the dispensing volume validation was not properly done and none of the compounders on the market can meet the stated specifications.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- The Metrix Company
- Distributor
- Product distributed to medical facilities in the following US states: AL, AZ, CA, CO, CT, FL, IA, IL, KS, LA, MA, MD, MN, NC, OH, PA, PR, TN, TX, UT, and WA. Product was also distributed internationally to the following countries: Australia, Canada, Dominican Republic, Malaysia, Turkey, and United Kingdom
- Lot
- All lots are being recalled. Lot code consists of the five digit product code followed by a dash followed by a unique five character numeric or alphanumeric sequence. Affected product codes for this product: 58740, 58741, 58742, 58749, 58750, 58753, 58754
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDASECURE 250 mL EMPTY EVA CONTAINER With 2 Ports, REF 66045; An empty single-use pouch made of plastic intendedNovember 20, 2024
- FDASECURE 100 mL EMPTY EVA CONTAINER With 2 Ports, REF 66040; An empty single-use pouch made of plastic intendedNovember 20, 2024
- FDAImed Products Imed EVA BAG, 150 mL, 2-PORTS, REF IM38040; An empty single-use pouch made of plastic intendedNovember 20, 2024
- FDACORMIX 150 mL EMPTY EVA CONTAINER With 2 Ports, REF MS64040; An empty single-use pouch made of plasticNovember 20, 2024
- FDAAUTOCOMP6 XP High Speed Compounder REF 58810August 3, 2016
- FDAAcceava Mono II Rapid Test Strip (Whole Blood, Serum, Plasma), Catalog No. 92404. Mononucleosis Rapid TestAugust 3, 2016
Questions about this recall
Was Container fill and transfer sets, automatic fill compatible, 250 mL, 500 mL, 1000 mL, 1500 mL, 2000 mL, 3000 mL, or 4000 mL. recalled?
Yes. FDA published recall Z-2261-2016 on August 3, 2016. Reported hazard: An issue was identified after a redesign of the auto compounding device. The compounder cannot meet the designed accuracy specification when dispensing into bags. After reviewing the initial design of the device, it was determined the dispensing volume validation was not properly done and none of the compounders on the market can meet the stated specifications.
What should I do if I have this product?
Voluntary: Firm initiated · Letter