Back to search
FDAFDA · Food and drugs·Recall number Z-2262-2013

QUADROX iD HMOD 2030 Diffusion Membrane Oxygenator; QUADROX-i ADULT HMO 70000 / HMO 71000; QUADROX-iD Pediatric HMOD 30000. The QUADROX oxygenators are intended for use in extracorporeal circulation during cardiopulmonary bypass in cardiac surgery.

Published October 2, 2013 · Recall initiated: August 15, 2013

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The manufacturer has conducted a review of the results of product testing performed during the last twelve months and has identified specific lots of product which may not meet specifications for bond strength on the inlet and/or outlet connector.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Maquet Cardiovascular Us Sales, Llc
Distributor
Worldwide Distribution, including Nationwide (US) and foreign countries.
Lot
QUADROX-iD Adult catalog number HMOD 70000, lot # 70081567 QUADROX-i Adult catalog number VKMO 70000, lot # 70082663 QUADROX-i Adult catalog number VKMO 71000, lot # 70083974
Risk classification
Class II
Area (FDA)
device
Country
United States

Sell products like this?

Recalert watches your whole catalog for recalls like this one, automatically.

Start free — no card