FDAFDA · Food and drugs·Recall number Z-2263-2013
Reliance Endoscope Processing System (REPS), STERIS Corporation, 5960 Heisley Road, Mentor, OH 44060-1834 USA www.steris.com. The intended use of the REPS is washing and high level disinfection of up to two manually pre-cleaned, immersible, reusable, heat-resistant, semi-critical devices.
Published October 2, 2013 · Recall initiated: August 29, 2013
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The air pipe that delivers unfiltered ambient air from the procedure room into the water bottle on the listed FUJINON G5 & G8 type flexible GI endoscopes does not connect to the main air/water channel within the light guide (LG) connection head and, as a result, the air pipe does not receive forced flow during the Reliance EPS processing cycle.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Steris Corporation
- Distributor
- Worldwide Distribution-USA (nationwide) including the states of AK, CO, FL, GA, HI, ID, IL, KS, LA, MD, MN, MO, MT, NE, NM, OH, PA, SC, UT, & WY and the countries of Australia, Canada & Spain.
- Lot
- Model #'s: MB30001, MB30002, MB30003, MB000077, MB000078, MB000079, MB000080 and MB000081. Serial number range 3625606001 - 3608912002
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States