FDAFDA · Food and drugs·Recall number Z-2267-2013
DRI Cocaine Metabolite Assay, labeled as Multigent Cocaine Assay for Abbott Laboratories. 100 ml, IVD Mfg by: Microgenics Corporation, 46360 Fremont Blvd. Fremont, CA 94538 The DRI Cocaine Metabolite Enzyme Immunoassay is a homogeneous enzyme intended for the qualitative and semi-quantitative determination of benzoylecgonine (Cocaine Metabolite) in human urine with either 300 ng/mL or 150 ng/mL as a cutoff calibrator. The assay provided only a preliminary analytical test result. A more specific alternative chemical method must be used to obtain a confirm analytical result.
Published October 2, 2013 · Recall initiated: July 23, 2013
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Surveillance of reagent stability for specific lots of DRI Cocaine Metabolite Reagent show an unexpected decreasing change in the reaction rate, indicating a faster than expected reagent degradation.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Microgenics Corp
- Distributor
- US Distribution including the states of DC, MA, NJ, TX. FL, CA, WA, NY, TX, CT, KY, LA, AR, TN, AL, GA, CT, VA, NC, KS, MD, OH and IL
- Lot
- Part no. 0055, lot number 60053405, Exp 2/29/2016; Part No. 0056, lot number 60053396, Exp 2/29/2016; Part No. 10014593, lot number 60112588, Exp 2/29/2016; - not distributed Part number #l40-20, lot number 24215M500, Exp 8/13/14.
- Risk classification
- Class II
- Area (FDA)
- device
- Country