Back to search
FDAFDA · Food and drugs·Recall number Z-2267-2024

IsoLoc Prostate Immobilization Treatment Balloon Device Kit, Catalog/UPN: ISO-100

Published July 10, 2024 · Recall initiated: May 21, 2024

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

A single batch of ISO-100 IsoLoc Gas-Release Endorectal Balloons is mislabeled. The affected ISO-100 balloon catheter has a nominal fill volume of 100cc/ml but the inner pouch identifies the product as an ISO-60 (fill volume of 60cc/ml).

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Angiodynamics, Inc.
Distributor
US Nationwide distribution in the states of AZ, FL, IL, MO, MS, NJ, NM, NY, PA, TX, WI.
Lot
UPN/Catalog: ISO-100; UDI/DI: 10851546007100 (Box) and 00851546007103 (Pouch); Lot number: 5402.
Risk classification
Class II
Area (FDA)
device
Country
United States