FDAFDA · Food and drugs·Recall number Z-2285-2026
Philips Azurion 3M15, Model Numbers: 722064, 722222 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key
Published June 10, 2026 · Recall initiated: December 7, 2023
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run.
What to do
Letter
Recall details
- Manufacturer
- PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
- Distributor
- US Nationwide distribution.
- Lot
- UDI (01)00884838085282(21)209, (01)00884838099210(21)100, (01)00884838099210(21)254, (01)00884838099210(21)256, (01)00884838099210(21)257, (01)00884838099210(21)265, (01)00884838099210(21)255, (01)00884838099210(21)99, (01)00884838085282(21)260, (01)00884838085282(21)261, (01)00884838099210(21)69, (01)00884838099210(21)12, (01)00884838085282(21)314, (01)00884838085282(21)259, (01)00884838099210(21)15, (01)00884838099210(21)19, (01)00884838099210(21)23, (01)00884838099210(21)17, (01)00884838085282(21)258, (01)00884838099210(21)55, (01)00884838099210(21)45, (01)00884838085282(21)270, (01)00884838099210(21)187, (01)00884838099210(21)223, (01)00884838099210(21)44, (01)00884838099210(21)113, (01)00884838099210(21)112.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- Netherlands