ABX Pentra Micro ALBUMIN 2 CP, Serum/Plasma, REF 1300032563 The device is intended for use in conjunction with certain materials to measure a variety of analytes.
Published July 4, 2018 · Recall initiated: April 26, 2018
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
When the ABX Pentra Micro ALBUMIN 2 CP (ALBT2) Reagent 2 is sampled and dispensed immediately after an ABX Pentra Creatinine 120 CP (CREA3) Reagent 2 dispense, the final result of the microalbumin may be elevated more than 10%. This increase may affect serum, plasma and urine sample types.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Horiba Instruments, Inc dba Horiba Medical
- Distributor
- Worldwide Distribution: US (nationwide) and to countries of: Bulgaria, Burma, Cameroon, Djibouti, France, Gabon, Germany, Italy, Lebanon, Mali, Mauritania, Poland, Portugal, Romania, Saudi Arabia, South Africa and Thailand.
- Lot
- All lot numbers; Software Version(s): Application versions below (V5.14 & V2.12) for Pentra 400 and Versions below (V1.07) for Pentra C400
- Risk classification
- Class II
- Area (FDA)
- device