FDAFDA · Food and drugs·Recall number Z-2291-2026
Philips Azurion 7M20, Model Numbers: 722079, 722224 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key
Published June 10, 2026 · Recall initiated: December 7, 2023
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run.
What to do
Letter
Recall details
- Manufacturer
- PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
- Distributor
- US Nationwide distribution.
- Lot
- UDI (01)00884838099258(21)450, (01)00884838099258(21)750, (01)00884838099258(21)409, (01)00884838099258(21)262, (01)00884838099258(21)275, (01)00884838099258(21)277, (01)00884838099258(21)295, (01)00884838099258(21)272, (01)00884838099258(21)261, (01)00884838099258(21)630, (01)00884838099258(21)1003, (01)00884838099258(21)1315, (01)00884838099258(21)381, (01)00884838099258(21)959, (01)00884838085268(21)1013, (01)00884838085268(21)1412, (01)00884838099258(21)232, (01)00884838085268(21)460, (01)00884838085268(21)1387, (01)00884838099258(21)260, (01)00884838099258(21)1019, (01)00884838099258(21)1371, (01)00884838085268(21)40, (01)00884838085268(21)43, (01)00884838085268(21)35, (01)00884838099258(21)1049, (01)00884838099258(21)1484, (01)00884838099258(21)1083, (01)00884838085268(21)703307, (01)00884838099258(21)246, (01)00884838099258(21)475.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- Netherlands