FDAFDA · Food and drugs·Recall number Z-2315-2024
iLED 7 surgical lights, Model Numbers: a) 4068110 (Ceiling Single Surgical Light); b) 4068120 (Mobile Surgical Light); c) 4068140 (Pendant Surgical Light); d) 4068210 (Ceiling Duo Surgical Lights); e) 4068310 (Ceiling Trio Surgical Lights); f) 4068410 (Ceiling Quad Surgical Lights)
Published July 17, 2024 · Recall initiated: June 4, 2024
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Baxter Healthcare Corporation is issuing an Urgent Medical Device Correction due to customer reports of thermal skin injuries when using the iLED 7 surgical light system products listed below. Specific VidiaPort and Trulight product codes are also impacted as they are equipped with iLED 7 light-heads
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Baxter Healthcare Corporation
- Distributor
- Worldwide distribution.
- Lot
- All Serial Numbers: a) UDI/DI 00887761968325; b) UDI/DI 00887761968318; c) UDI/DI 00887761968301; d) UDI/DI 00887761968295; e) UDI/DI 00887761968288; f) UDI/DI 00887761968271
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States