GE SIGNA Hero XT (China Only), Nuclear Magnetic Resonance Imaging System

Published August 20, 2025 · Recall initiated: June 20, 2025

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA Artist Evo, SIGNA Voyager AIR (China Only), SIGNA Architect AIR (China Only), SIGNA Premier XT (China Only), SIGNA Premier Elite (China Only), SIGNA Premier MAX (China Only), SIGNA Hero XT (China Only), SIGNA Hero Elite (China Only), SIGNA Hero MAX (China Only), SIGNA PET/MR AIR (China Only), SIGNA PET/MR AIR Plus (China Only), the planned maintenance steps to check the functionality of the caster locks on the MR patient table may not have been performed.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
GE Medical Systems, LLC
Distributor
Worldwide distribution.
Lot
GTIN: Unknown; All System ID numbers that GE HealthCare performed planned maintenance on since March 1, 2023
Risk classification
Class II
Area (FDA)
device
Country
United States