VersaTREK Windows Software, Version 5.4.3 is a component of the VersaTrek System. The Software CD will be individually packaged using a standard CD plastic case and placed into a standard CD mailer The VersaTREK System is for cultivating and recovering microorganisms, especially bacteria and yeasts, from blood and other normally sterile body fluids.

Published September 12, 2012 · Recall initiated: July 23, 2012

Summary

FDA records medical device Class III recall Z-2323-2012, published September 12, 2012 and initiated July 23, 2012. Recalling firm: Trek Diagnostic Systems; Status: Terminated; Reported quantity: 333 units; Lot: Software: Cat. 6133-30-3, Version 5.4.3. Class III means the product is not likely to cause adverse health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Review of instrument software databases and investigation of two customer reports determined that on rare occasions (less than 0.3%), following the removal of a positive signaling bottle from the VersaTREK instrument, the positive test result inadvertently reverted to a negative result in the instrument database.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Trek Diagnostic Systems
Distributor
Worldwide Distribution - USA including AL, AR, AZ, CA, CO , CT, FL, GA, IA, ID, IL, IN, KS, LA, MA, MD, ME, MI, MN, MO, NC, NE, NJ, NY, OH, OK, OR, PA, RI, SD, TN, TX, VA, VT, WA, WI, WV, and WY. Internationally to the countries of Bahamas, Bangladesh, Brazil, Canada, Chile, China, Colombia, Czech Republic, France, India, Ireland, Italy, Japan, Jordan, Kenya, Kuwait, Malaysia, Mexico, Pakistan, Peru, Philippines, Portugal, Puerto Rico, Russian Federation, South Africa, Spain, Thailand, Turkey, and United Kingdom
Lot
Software: Cat. 6133-30-3, Version 5.4.3.
Risk classification
Class III
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was VersaTREK Windows Software, Version 5.4.3 is a component of the VersaTrek System. The Software CD will be individually packaged using a standard CD plastic case and placed into a standard CD mailer The VersaTREK System is for cultivating and recovering microorganisms, especially bacteria and yeasts, from blood and other normally sterile body fluids. recalled?

Yes. FDA published recall Z-2323-2012 on September 12, 2012. Reported hazard: Review of instrument software databases and investigation of two customer reports determined that on rare occasions (less than 0.3%), following the removal of a positive signaling bottle from the VersaTREK instrument, the positive test result inadvertently reverted to a negative result in the instrument database.

What should I do if I have this product?

Voluntary: Firm initiated · Letter