VersaTREK Windows Software, Version 5.4.3 is a component of the VersaTrek System. The Software CD will be individually packaged using a standard CD plastic case and placed into a standard CD mailer The VersaTREK System is for cultivating and recovering microorganisms, especially bacteria and yeasts, from blood and other normally sterile body fluids.
Published September 12, 2012 · Recall initiated: July 23, 2012
Summary
FDA records medical device Class III recall Z-2323-2012, published September 12, 2012 and initiated July 23, 2012. Recalling firm: Trek Diagnostic Systems; Status: Terminated; Reported quantity: 333 units; Lot: Software: Cat. 6133-30-3, Version 5.4.3. Class III means the product is not likely to cause adverse health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Review of instrument software databases and investigation of two customer reports determined that on rare occasions (less than 0.3%), following the removal of a positive signaling bottle from the VersaTREK instrument, the positive test result inadvertently reverted to a negative result in the instrument database.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Trek Diagnostic Systems
- Distributor
- Worldwide Distribution - USA including AL, AR, AZ, CA, CO , CT, FL, GA, IA, ID, IL, IN, KS, LA, MA, MD, ME, MI, MN, MO, NC, NE, NJ, NY, OH, OK, OR, PA, RI, SD, TN, TX, VA, VT, WA, WI, WV, and WY. Internationally to the countries of Bahamas, Bangladesh, Brazil, Canada, Chile, China, Colombia, Czech Republic, France, India, Ireland, Italy, Japan, Jordan, Kenya, Kuwait, Malaysia, Mexico, Pakistan, Peru, Philippines, Portugal, Puerto Rico, Russian Federation, South Africa, Spain, Thailand, Turkey, and United Kingdom
- Lot
- Software: Cat. 6133-30-3, Version 5.4.3.
- Risk classification
- Class III
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDASiemens Mobilett Mira. Mobile X-Ray System.September 12, 2012
- FDABrand name: RayStation. Generic name: Treatment Planning System. Software version and build numbers areSeptember 12, 2012
- FDAAmeriWater MRO Dialysis RO+ System, AmeriWater 1303 Stanley Avenue, Dayton, OH 45404 (800) 535-5585September 12, 2012
- FDA100 ft roll of tubing inside a cardboard box. Medical devices are sealed into the tubes prior to ethyleneSeptember 12, 2012
- FDABD GeneOhm VanR Assay , Catalog # 441250, 5 box tests labeled in part***GeneOhm Sciences Canada, Inc., 2555September 12, 2012
- FDAPolar Dressing, Knee/MU/Shldr Intelli-Flo-Breg, PN 10630 Used with cold therapy pads to provide insulationSeptember 12, 2012
Questions about this recall
Was VersaTREK Windows Software, Version 5.4.3 is a component of the VersaTrek System. The Software CD will be individually packaged using a standard CD plastic case and placed into a standard CD mailer The VersaTREK System is for cultivating and recovering microorganisms, especially bacteria and yeasts, from blood and other normally sterile body fluids. recalled?
Yes. FDA published recall Z-2323-2012 on September 12, 2012. Reported hazard: Review of instrument software databases and investigation of two customer reports determined that on rare occasions (less than 0.3%), following the removal of a positive signaling bottle from the VersaTREK instrument, the positive test result inadvertently reverted to a negative result in the instrument database.
What should I do if I have this product?
Voluntary: Firm initiated · Letter