Decompression Table Model E9011. This device applies cervical and/or lumbar decompression with practitioner definable target treatment location, treatment pounds, and patterns of treatment.
Published July 17, 2024 · Recall initiated: April 29, 2024
Summary
FDA records medical device Class II recall Z-2326-2024, published July 17, 2024 and initiated April 29, 2024. Recalling firm: Pivotal Health Solutions, Inc.; Status: Ongoing; Reported quantity: 32 tables; Lot: UDI-DI 00850008449826; Model Number: E9011; Serial Numbers: WTN-DOC03132457209 WTN-DOC-3132457210, WTN-DOC03132457235, WTN-DOC03132457237, WTN-DOC03132457238, WTN-DOC03132457239 Expanded Recall serial numbers: WTN-DOC04262458492, WTN-DOC04262458490, WTN-DOC04262458487, WTN-DOC04262458488, WTN-DOC04262458489, WTN-DOC0332457239, WTN-DOC04082457899, WTN-DOC04262458484, WTN-DOC03132457238, WTN-DOC03132457237, WTN-DOC03132457210, WTN-DOC04082457897, WTN-DOC04082457898, WTN-DOC03132457201, WTN-DOC04192458307, WTN-DOC04082457896, WTN-DOC03132457209, WTN-DOC03132457235, WTN-DOC04262458496, WTN-DOC03132457240, WTN-DOC03132457241, WTN-DOC04082457882, WTN-DOC04172458185, WTN-DOC04082457893, WTN-DOC04082457894, WTN-DOC04082457895. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
This product is being recalled due to potentially insufficient rework on a part (E7130-B) on the lumbar section of the frame on 6 distributed tables. There were 4 additional tables with the faulty E7130-B part that were manufactured, but had not been distributed yet.
What to do
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Pivotal Health Solutions, Inc.
- Distributor
- Distributed to CA, FL, TX, ID, GA, and TN
- Lot
- UDI-DI 00850008449826; Model Number: E9011; Serial Numbers: WTN-DOC03132457209 WTN-DOC-3132457210, WTN-DOC03132457235, WTN-DOC03132457237, WTN-DOC03132457238, WTN-DOC03132457239 Expanded Recall serial numbers: WTN-DOC04262458492, WTN-DOC04262458490, WTN-DOC04262458487, WTN-DOC04262458488, WTN-DOC04262458489, WTN-DOC0332457239, WTN-DOC04082457899, WTN-DOC04262458484, WTN-DOC03132457238, WTN-DOC03132457237, WTN-DOC03132457210, WTN-DOC04082457897, WTN-DOC04082457898, WTN-DOC03132457201, WTN-DOC04192458307, WTN-DOC04082457896, WTN-DOC03132457209, WTN-DOC03132457235, WTN-DOC04262458496, WTN-DOC03132457240, WTN-DOC03132457241, WTN-DOC04082457882, WTN-DOC04172458185, WTN-DOC04082457893, WTN-DOC04082457894, WTN-DOC04082457895
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAKosmos on iOS impacting iOS, Model Numbers: P007639, P007944, software is part of a general-purpose diagnosticJuly 17, 2024
- FDAPresource kits and trays, labeled as: 1) ANGIOGRAPHY PACK, Catalog Number SAN13BS605; 2) ANGIOGRAPHY PACK,July 17, 2024
- FDAZOLL Powerheart G5 AED, Semi-Automatic, G5Sxxx Family -Automated external defibrillator (AED) is designed forJuly 17, 2024
- FDAPresource kits and trays, labeled as: 1) Cardiac Catheterization Kit, Catalog Number SAN69CCKRB; 2) CardiacJuly 17, 2024
- FDAWelch Allyn CP150 Electrocardiograph with Spirometry Option, Product Code: CP150A-1ENB. The ElectrocardiographJuly 17, 2024
- FDAPresource kits and trays, labeled as: 1) Cardiovascular Procedure Kit, Catalog Number SAN29ANCC3; 2)July 17, 2024
Questions about this recall
Was Decompression Table Model E9011. This device applies cervical and/or lumbar decompression with practitioner definable target treatment location, treatment pounds, and patterns of treatment. recalled?
Yes. FDA published recall Z-2326-2024 on July 17, 2024. Reported hazard: This product is being recalled due to potentially insufficient rework on a part (E7130-B) on the lumbar section of the frame on 6 distributed tables. There were 4 additional tables with the faulty E7130-B part that were manufactured, but had not been distributed yet.
What should I do if I have this product?
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit