microTargeting" STar" Drive System (Manual), Product Number ST-DS-MA

Published July 11, 2018 · Recall initiated: May 10, 2018

Summary

FDA records medical device Class II recall Z-2349-2018, published July 11, 2018 and initiated May 10, 2018. Recalling firm: FHC, Inc.; Status: Terminated; Reported quantity: 2952 total; Lot: UDI: 00873263004418. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The FHC Depth Stop Adapter may cause damage to the lead and stylet when the depth stop screw is over-tightened onto the lead. This damage can manifest as low impedances or a short circuit between one or multiple electrode pairs in the lead, and has resulted in the need to remove and replace the implanted lead intraoperative or in a follow-up procedure.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
FHC, Inc.
Distributor
The devices were distributed to the following US states: AL, AR, AZ, CA, CO, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VT, WA, WI, WV, and WY. The devices were distributed to the following foreign countries: Argentina, Australia, Canada, Columbia, Denmark, Ecuador, France, Germany, Great Britain, Italy, Japan, Norway, Romania, South Africa, Switzerland, Saudi Arabia, Taiwan.
Lot
UDI: 00873263004418
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was microTargeting" STar" Drive System (Manual), Product Number ST-DS-MA recalled?

Yes. FDA published recall Z-2349-2018 on July 11, 2018. Reported hazard: The FHC Depth Stop Adapter may cause damage to the lead and stylet when the depth stop screw is over-tightened onto the lead. This damage can manifest as low impedances or a short circuit between one or multiple electrode pairs in the lead, and has resulted in the need to remove and replace the implanted lead intraoperative or in a follow-up procedure.

What should I do if I have this product?

Voluntary: Firm initiated · Letter