Thoratec Heartmate 3 Sealed Outflow Graft with Bend Relief, Sterile, REF 105581US/105581, RX only, Part of Left Ventricular Assist Device (LVAD) UDI: 00813024013266
Published August 28, 2019 · Recall initiated: March 30, 2019
Summary
FDA records medical device Class II recall Z-2349-2019, published August 28, 2019 and initiated March 30, 2019. Recalling firm: Thoratec Corp.; Status: Terminated; Reported quantity: 149 outflow graft units; Lot: 105581/105581US: The shelf life is 60 months from date of manufacture. All batch/lot numbers 6613000 and below or 50000000 and higher. UDI/GMDN 00813024013266 / 61747. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Outflow Graft leaking at the pump connection during the implantation process. If the C-Ring is insufficiently seated, it renders the device nonfunctional because the Outflow Graft connection will leak during the implant procedure.
What to do
Voluntary: Firm initiated
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Thoratec Corp.
- Distributor
- Worldwide distribution. US nationwide, Australia, Austria, Brazil, Canada, Cayman Islands, Chile, Colombia, Egypt, France, Germany, Greece, Singapore, Iran, Italy, Kuwait, Lebanon, Romania, Switzerland, Thailand, United Kingdom.
- Lot
- 105581/105581US: The shelf life is 60 months from date of manufacture. All batch/lot numbers 6613000 and below or 50000000 and higher. UDI/GMDN 00813024013266 / 61747
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAThoratec HeartMate System Monitor, REF: 1286, L1286, 1286A, L1286A, 1286A-US, L1286A-US, 1286C, L1286C, L1386,June 19, 2024
- FDAThoratec HeartMate 3, LVAS Implant Kit, REF:106524US, Sterile EO, RxOnlyMay 8, 2024
- FDAThoratec HeartMate 3, LVAS Implant Kit, REF:106524INT, Sterile EO, RxOnlyMay 8, 2024
- FDAHeartMate 3 Left Ventricular Assist System (LVAS): 1. HeartMate 3 Sealed Outflow Graft with Bend Relief,March 27, 2024
- FDAKit BD Max Enteric Viral Panel RUO; Catalog # 443715August 28, 2019
- FDAACUSON NX2 Elite Diagnostic Ultrasound System, Model # 11284500, RX Only, Siemens Medical Solutions,August 28, 2019
Questions about this recall
Was Thoratec Heartmate 3 Sealed Outflow Graft with Bend Relief, Sterile, REF 105581US/105581, RX only, Part of Left Ventricular Assist Device (LVAD) UDI: 00813024013266 recalled?
Yes. FDA published recall Z-2349-2019 on August 28, 2019. Reported hazard: Outflow Graft leaking at the pump connection during the implantation process. If the C-Ring is insufficiently seated, it renders the device nonfunctional because the Outflow Graft connection will leak during the implant procedure.
What should I do if I have this product?
Voluntary: Firm initiated