MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 20 cm, Straight Extensions, IC Tray REF: 8888102003HP
Published August 23, 2023 · Recall initiated: June 28, 2023
Summary
FDA records medical device Class I recall Z-2371-2023, published August 23, 2023 and initiated June 28, 2023. Recalling firm: Covidien LP; Status: Ongoing; Reported quantity: 1875 units; Lot: GTIN: 20884521127996 LOT Numbers: 2204000239 2227800110. Class I means a reasonable probability of serious adverse health consequences or death.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Catheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or excessive MDX may dislodge from the catheter lead to full catheter obstruction resulting in delay to treatment and partial obstruction resulting in reduced flow, or particulate dislodgement that may result in delay to treatment, hemolysis, embolism/embolus or thrombosis/thrombus
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Covidien LP
- Distributor
- US Nationwide - Worldwide Distribution Foreign: Argentina Austria Bahrain Bangladesh Belgium Brazil Brunei Darussalam Canada Cayman Islands China Costa Rica Czech Republic Denmark Dominican Republic Finland France French Polynesia Germany Greece Hungary India Iran, Islamic Republic Of Ireland Italy Kuwait Malaysia Maldives Mauritius Mexico Nepal Netherlands New Caledonia Nicaragua Norway Philippines Poland Portugal Puerto Rico Saudi Arabia Singapore Slovakia South Africa Spain Sri Lanka Sweden Switzerland Thailand Turkey United Arab Emirates United Kingdom
- Lot
- GTIN: 20884521127996 LOT Numbers: 2204000239 2227800110
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAChameleon PTA Balloon Catheter with Injection Port 5mm x 40mm x 75cm (CFN CH05-40-75US), 6mm x 40mm x 75cmNovember 6, 2024
- FDAPalindrome Precision HSI Chronic Catheter Sport Pack 14.5 Fr/Ch (4.8mm) x 19 cm (Symmetrical Tip, HeparinMay 1, 2024
- FDAPalindrome Precision H Chronic Catheter Kit 14.5 Fr/Ch (4.8mm) x 23 cm (Symmetrical Tip, Heparin Coating andMay 1, 2024
- FDAAbsorbaTack Absorbable Fixation Device 30 Violet Absorbable Tacks 5 mm- -intended for fixation of prostheticFebruary 14, 2024
- FDATim Heung Yuen Black Melon Seed 6 oz., 12 oz. Ingredients: Melon Seeds, Salt, Sugar & Anise Seed DistributedAugust 23, 2023
- FDAMAHURKAR 12 Fr Triple Lumen Acute Dialysis Catheter, 13 cm, Curved Extensions, Kit" REF: 8888345603August 23, 2023
Questions about this recall
Was MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 20 cm, Straight Extensions, IC Tray REF: 8888102003HP recalled?
Yes. FDA published recall Z-2371-2023 on August 23, 2023. Reported hazard: Catheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or excessive MDX may dislodge from the catheter lead to full catheter obstruction resulting in delay to treatment and partial obstruction resulting in reduced flow, or particulate dislodgement that may result in delay to treatment, hemolysis, embolism/embolus or thrombosis/thrombus
What should I do if I have this product?
Voluntary: Firm initiated · Letter