FDAFDA · Food and drugs·Recall number Z-2379-2012
HALO Flex Energy Generators Manufactured by B¿RRX Medical (now Covidien) 540 Oakmead Parkway Sunnyvale, CA 95085. The HALO Flex Energy Generator is indicated for use for the coagulation of soft tissue. The HALO Flex System is indicated for use in the coagulation of bleeding and non-bleeding sites in the gastrointestinal tract including but not limited to, the esophagus, including Barrett's Esophagus.
Published September 26, 2012 · Recall initiated: August 3, 2012
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Potential failure of HALO Energy Generator to enter "stand by" mode when initially powered on; and therefore can not be used for patient care.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Covidien LLC
- Distributor
- Worldwide Distribution - USA (nationwide) and the countries of Argentina, Australia, Austria, Belgium, Canada, Chile, Colombia, Czech Republic, Denmark, France, Germany, Greece, Hungary, India, Israel, Jordan, Netherlands, Norway, Poland, Portugal, Saudi Arabia, South Africa, Spain, Sweden, Switzerland, Taiwan, Venezuela and UK.
- Lot
- Model numbers 1190A-115A and 1190A-230A; all serial numbers.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States