Baxter prismaflex, HF 1000 PAES membrane, REF 107140 Used with Prismaflex or PrisMax control units for continuous fluid management and renal replacement therapies.

Published September 8, 2021 · Recall initiated: July 22, 2021

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Products labeled as sterile were distributed, but may not have been sterilized. Baxter confirmed Steril Milano observed nonconformances that were not reported to Baxter Data was manipulated to appear to be conforming

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Baxter Healthcare Corporation
Distributor
US Nationwide distribution.
Lot
Lot Numbers: 20B2330M
Risk classification
Class II
Area (FDA)
device
Country
United States