Brand Name: Atlantis IO FLO Product Name: Atlantis IO FLO-S Kit Model/Catalog Number: A04S / 6802034 Product Description: The Atlantis intraoral feature locating object for suprastructures (IO FLO-S) is intended to support the prosthetic procedure with the aim of restoring chewing function in partial or complete edentulous patients, by enabling digital impression taking on implant or abutment level for all positions in the mouth of the patient or on the master model. Component: not a component
Published September 3, 2025 · Recall initiated: July 24, 2025
Summary
FDA records medical device Class II recall Z-2422-2025, published September 3, 2025 and initiated July 24, 2025. Recalling firm: Dentsply IH, Inc.; Status: Ongoing; Reported quantity: 780 units; Lot: Catalog / Ref #: 68020034; UDI: 07392532249440; Lot #s: 10515863, 10512084, 10505543, 10556741. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The defect in the thread area will not allow the device to fully engage with an implant or analog (implant replica used for prosthetic modeling). The threads will only engage approximately 2 rotations as opposed to full engagement which is approximately 5 rotations. This equates to approximately 1mm difference in depth of engagement.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Dentsply IH, Inc.
- Distributor
- Worldwide - US Nationwide distribution in the states of IL, LA, MA, MN, MO, ND, NJ, NM, OH, OK, PA, TX, WV and the countries of AT, AU, BE, CA, DE, DK, ES, FI, FR, IT, JP, NL, NO, PL, QA, SE, UA, UK.
- Lot
- Catalog / Ref #: 68020034; UDI: 07392532249440; Lot #s: 10515863, 10512084, 10505543, 10556741;
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDACorin Operating Tables Model Numbers: (1) 770001A0, (2) 770001A2, (3) 770001B0, (4) 770001B2, (5) 770001F0,September 3, 2025
- FDAPrisMax V2-US, Hemodialysis Delivery System, Product Codes: 955626, 955558, 955627, 955701September 3, 2025
- FDABrand Name: THOR Product Name: NovoTHOR Gen 3.0 Reg PRX Model/Catalog Number: S2186 Software Version: N/ASeptember 3, 2025
- FDAOrascoptic Superior Visualization Dragonfly Essentials Kit ( Nose pad, a screw, lens cleaning wipe, sideSeptember 3, 2025
- FDAX5-1 Transducer ProbeSeptember 3, 2025
- FDAOrascoptic Superior Visualization Phantom Nose Pad Accessory Kit ( nose pad and screw) 1 pk. and 20 pk.September 3, 2025
Questions about this recall
Was Brand Name: Atlantis IO FLO Product Name: Atlantis IO FLO-S Kit Model/Catalog Number: A04S / 6802034 Product Description: The Atlantis intraoral feature locating object for suprastructures (IO FLO-S) is intended to support the prosthetic procedure with the aim of restoring chewing function in partial or complete edentulous patients, by enabling digital impression taking on implant or abutment level for all positions in the mouth of the patient or on the master model. Component: not a component recalled?
Yes. FDA published recall Z-2422-2025 on September 3, 2025. Reported hazard: The defect in the thread area will not allow the device to fully engage with an implant or analog (implant replica used for prosthetic modeling). The threads will only engage approximately 2 rotations as opposed to full engagement which is approximately 5 rotations. This equates to approximately 1mm difference in depth of engagement.
What should I do if I have this product?
Voluntary: Firm initiated · Letter