WATCHMAN FXD Curve Access System Dbl, US, Material Number (UPN) M635TU80020; intended to provide vascular and transseptal access for the family of WATCHMAN FLX LAAC Devices with Delivery Systems.

Published September 10, 2025 · Recall initiated: July 29, 2025

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Boston Scientific has identified a higher likelihood of air embolism events when procedures are performed without positive pressure-controlled ventilation. According to published literature and clinical data, in percutaneous procedures requiring transseptal access to the left atrium when conscious or deep sedation is used, patients have an approximately three-times higher risk (U.S. study) of negative left atrium pressure and air ingress. This risk is especially prevalent in patients with pre-existing low left atrial pressure, hypovolemia, and partial upper airway collapse.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Boston Scientific Corporation
Distributor
Worldwide.
Lot
GTIN 00191506013813, ALL NON-EXPIRED BATCHES
Risk classification
Class I
Area (FDA)
device
Country
United States