FDAFDA · Food and drugs·Recall number Z-2439-2019

BD Kiestra InoqulA+ Software Version 20.3. IVD automated specimen processing Catalog Number : 447213

Published September 11, 2019 · Recall initiated: July 26, 2019

Summary

FDA records medical device Class II recall Z-2439-2019, published September 11, 2019 and initiated July 26, 2019. Recalling firm: Becton Dickinson & Co.; Status: Terminated; Reported quantity: 9 units; Lot: Catalog Number: 447213 Serial Number/UDI: INO-000290 00382904472139 INO-000263 00382904472139 INO-000264 00382904472139 20130276 00382904472139 INO-000282 00382904472139 INO-000284 00382904472139 INO-000272 00382904472139 INO-000274 00382904472139. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

An anomaly is present in InoqulA / InoqulA+ software version 20.3, has the potential to cause a mismatch between a specimen and plate

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Becton Dickinson & Co.
Distributor
Illinois, Iowa, Missouri, New York, Ohio, Tennessee World Wide (5): Singapore, Taiwan, France, Netherlands
Lot
Catalog Number: 447213 Serial Number/UDI: INO-000290 00382904472139 INO-000263 00382904472139 INO-000264 00382904472139 20130276 00382904472139 INO-000282 00382904472139 INO-000284 00382904472139 INO-000272 00382904472139 INO-000274 00382904472139
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was BD Kiestra InoqulA+ Software Version 20.3. IVD automated specimen processing Catalog Number : 447213 recalled?

Yes. FDA published recall Z-2439-2019 on September 11, 2019. Reported hazard: An anomaly is present in InoqulA / InoqulA+ software version 20.3, has the potential to cause a mismatch between a specimen and plate

What should I do if I have this product?

Voluntary: Firm initiated · Letter