Back to search
FDAFDA · Food and drugs·Recall number Z-2440-2024

Sensis Vibe systems with software version VD15B in combination with AXIOM Sensis Post Processing Workstation-Intended to be used as a diagnostic and administrative tool supporting hemodynamic catheterizations and/or electrophysiology studies, for cardiac as well as interventional Radiology as well as surgical studies. Equipped by modules, enabling various configurations. Model: 6648161

Published July 31, 2024 · Recall initiated: July 3, 2024

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

If the Sensis documentation functionality is used during adding of once-per-study reporting events (e.g., Type 1 events, as further defined in the administrator manual), the possibility exists that the application could crash.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Siemens Medical Solutions USA, Inc
Distributor
US Nationwide distribution in the states of CA, CT, FL, IA, KS, MA, MI, NJ, TX, VA.
Lot
UDI-DI: N/A Serial Numbers: 18904 18906
Risk classification
Class II
Area (FDA)
device
Country
United States

Sell products like this?

Recalert watches your whole catalog for recalls like this one, automatically.

Start free — no card