EMBLEM S-ICD Pulse Generator Model A209; Subcutaneous Implantable Cardioverter Defibrillator

Published August 30, 2023 · Recall initiated: July 11, 2023

Summary

FDA records medical device Class II recall Z-2441-2023, published August 30, 2023 and initiated July 11, 2023. Recalling firm: Boston Scientific Corporation; Status: Ongoing; Reported quantity: 3,856 units; Lot: GTIN: 00802526544101, 00802526548406, 00802526575105, 00802526575112, 00802526575129, 00802526575136, 00802526575143, 00802526575167, 00802526575181, 00802526575204, 00802526575211, 00802526575228, 00802526599002; All EMBLEM S-ICDs enrolled in LATITUDE. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

There is a potential for a rare interaction between the EMBLEM S-ICD and LATITUDE communicator, which may cause S-ICD sensing disablement for a 24-hour interval.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Boston Scientific Corporation
Distributor
Worldwide
Lot
GTIN: 00802526544101, 00802526548406, 00802526575105, 00802526575112, 00802526575129, 00802526575136, 00802526575143, 00802526575167, 00802526575181, 00802526575204, 00802526575211, 00802526575228, 00802526599002; All EMBLEM S-ICDs enrolled in LATITUDE
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was EMBLEM S-ICD Pulse Generator Model A209; Subcutaneous Implantable Cardioverter Defibrillator recalled?

Yes. FDA published recall Z-2441-2023 on August 30, 2023. Reported hazard: There is a potential for a rare interaction between the EMBLEM S-ICD and LATITUDE communicator, which may cause S-ICD sensing disablement for a 24-hour interval.

What should I do if I have this product?

Voluntary: Firm initiated · Letter