FDAFDA · Food and drugs·Recall number Z-2450-2015

Dimension Vista B¿2-Microglobulin Flex Reagent Cartridge (B2MIC); Product Usage: The B2MIC method is an in vitro diagnostic test for the quantitative measurement of B¿2-Microglobulin in human serum, heparinized plasma, EDTA plasma and urine on the Dimension Vista System. Measurements of B¿2-Microglobulin aid in the diagnosis of active rheumatoid arthritis and kidney disease.

Published September 2, 2015 · Recall initiated: July 23, 2015

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

customers are receiving a high number of Abnormal Assay Errors or are not able to calibrate B2MIC when using Flex reagent cartridge B2MIC various lots. In failure mode, there is the potential for falsely depressed beta2-Microglobulin results on the Dimension Vista system due to assay reagent contamination.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Siemens Healthcare Diagnostics, Inc.
Distributor
US Nationwide Distribution in the states of TN, AZ, CT, Fl, CA, NY and WI
Lot
Lot Number:15037MA; ExpDate: 2016-07-14 Lot # 15175MA; ExpDate: 2015-11-17; Lot # 15204MA; Expdate: 2017-01-13; Lot #15246MA; expdate: 2017-02-26 Lot # 15267MA; expdate: 2017-02-26
Risk classification
Class II
Area (FDA)
device
Country
United States