Hologic Fluoroscan Mini C-Arm Model with InSight- FD Software Version 5.0.1and 5.0.2 The device is a Mini C-arm Fluoroscopic Imaging System designed to provide -physicians with general fluoroscopic visualization of a patient, including but not limited to surgical orthopedic and podiatry use, critical and emergency procedures, and light anatomy imaging situation.
Published October 3, 2012 · Recall initiated: September 14, 2012
Summary
FDA records medical device Class II recall Z-2453-2012, published October 3, 2012 and initiated September 14, 2012. Recalling firm: Hologic, Inc.; Status: Terminated; Reported quantity: 46 units; Lot: Software version 5.0.1and 5.0.2. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Fluoroscan Mini C-Arm models -InSight2, version 4.0 and 4.0.1 - DICOM images stored in PACS may be flipped 180®around vertical axis. InSight FD, version 5.0.1 and 5.0.2 - DICOM images stored in PACS may be lipped 180® around vertical axis as well as 180® around a horizontal axis.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Hologic, Inc.
- Distributor
- Worldwide Distribution - USA (nationwide) and the countries of Argentina, Australia,Belgium, Canada, China, Germany, Ecuador, Spain, France, Great Britain, Greece, Hong Kong, Italy, Japan Korea, Lithuvania, Neherlands, qatar, Russia, Saudi Arabia, Sweden, Singapore, Slovenia, Thailand, Taiwan, South Africa, and Switzerland.
- Lot
- Software version 5.0.1and 5.0.2
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAThe Rapid fFN Specimen Collection Tube (PN: 71550-001) is a component of the Rapid fFN Specimen Collection kitFebruary 12, 2025
- FDAPanther Fusion SARS-CoV-2/Flu A/B/RSV Assay, REF: PRD-07400January 1, 2025
- FDAAptima CMV Quant Assay, Reference numbers: AW-22600-001, AW-25509-001, and AW-26381-001 Amplification Reagent,April 3, 2024
- FDAPANTHER FUSION System REF 9121010000 The Panther System is an integrated nucleic acid testing system whichJanuary 31, 2024
- FDAHologic Fluoroscan Mini C-Arm Model with InSight2, Software version 4.0 and 4.0.1 The device is a Mini C-armOctober 3, 2012
- FDAProgel Pleural Air Leak Sealant, Model #: PGPS002 Lot #: 111202-002 Product Usage: The Progel Pleural Air LeakOctober 3, 2012
Questions about this recall
Was Hologic Fluoroscan Mini C-Arm Model with InSight- FD Software Version 5.0.1and 5.0.2 The device is a Mini C-arm Fluoroscopic Imaging System designed to provide -physicians with general fluoroscopic visualization of a patient, including but not limited to surgical orthopedic and podiatry use, critical and emergency procedures, and light anatomy imaging situation. recalled?
Yes. FDA published recall Z-2453-2012 on October 3, 2012. Reported hazard: Fluoroscan Mini C-Arm models -InSight2, version 4.0 and 4.0.1 - DICOM images stored in PACS may be flipped 180®around vertical axis. InSight FD, version 5.0.1 and 5.0.2 - DICOM images stored in PACS may be lipped 180® around vertical axis as well as 180® around a horizontal axis.
What should I do if I have this product?
Voluntary: Firm initiated · Letter