FDAFDA · Food and drugs·Recall number Z-2466-2012

BariAir Therapy System The BariAir Therapy System is a risk management treatment system designed for large and / or difficult-to-manage patients. The BariAir Therapy System provides low air-loss pressure management therapy, pulsation, percussion, Turn Assist. flexible patient positioning capabilities and built-in scales for patients weighing 300 Ib* to 850 Ib ([136*-386 kg] including accessories). The BariAir Therapy System is indicated to aid in: " Patients whose body weight and size pose a significant risk or care management issue to the patient or staff during the performance of routine nursing care. " Large patients weighing between 300 Ib* and 850 Ib (136*-386 kg) including accessories. " Large patients who are difficult to turn. " Preventing and treating pressure ulcers. " Large patients requiring percussion therapy.

Published October 3, 2012 · Recall initiated: September 17, 2012

Summary

FDA records medical device Class II recall Z-2466-2012, published October 3, 2012 and initiated September 17, 2012. Recalling firm: KCI USA, Inc.; Status: Terminated; Reported quantity: 575; Lot: Multiple Serial Numbers with Model Numbers: 404000,405500, 405500RF. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

KCI has issued a medical device correction for the BariAir Therapy System Hand Pendent. The BariAir Therapy System is not designed to allow the patient to place the bed into the patient exit position. KCI has received complaints indicating that patients used the hand pendent to inadvertently place the BariAir Therapy System into a pseudo-exit position. In a small number of instances, this allow

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
KCI USA, Inc.
Distributor
Worldwide Distribution - USA (nationwide) and the countries of Saudi Arabia, United Kingdom, Singapore, Germany, Switzerland, Kuwait, United Arabian Emirates, and Austria.
Lot
Multiple Serial Numbers with Model Numbers: 404000,405500, 405500RF
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was BariAir Therapy System The BariAir Therapy System is a risk management treatment system designed for large and / or difficult-to-manage patients. The BariAir Therapy System provides low air-loss pressure management therapy, pulsation, percussion, Turn Assist. flexible patient positioning capabilities and built-in scales for patients weighing 300 Ib* to 850 Ib ([136*-386 kg] including accessories). The BariAir Therapy System is indicated to aid in: " Patients whose body weight and size pose a significant risk or care management issue to the patient or staff during the performance of routine nursing care. " Large patients weighing between 300 Ib* and 850 Ib (136*-386 kg) including accessories. " Large patients who are difficult to turn. " Preventing and treating pressure ulcers. " Large patients requiring percussion therapy. recalled?

Yes. FDA published recall Z-2466-2012 on October 3, 2012. Reported hazard: KCI has issued a medical device correction for the BariAir Therapy System Hand Pendent. The BariAir Therapy System is not designed to allow the patient to place the bed into the patient exit position. KCI has received complaints indicating that patients used the hand pendent to inadvertently place the BariAir Therapy System into a pseudo-exit position. In a small number of instances, this allow

What should I do if I have this product?

Voluntary: Firm initiated · Letter