This voluntary medical device correction is related to Alcon products that present a 23G nonvalved trocar cannula and infusion cannula together. These items are available as components within Accurus 23G TotalPLUS Pak, Accurus 23G Combined Procedure Packs (0.9mm and 1.1mm), Constellation 23G Total PLUS Pak, Constellation 23G Combined Pack (0.9 mm and 1.1 mm) and 23G Small Parts Kit. Single use 23G non-valved trocar cannulas are used to provide internal access to the ocular cavity during ophthalmic surgery. Single use infusion cannulas are used for the infusion of fluid or air during ophthalmic surgery to maintain intraocular pressure during posterior segment surgeries.
Published October 3, 2012 · Recall initiated: May 29, 2012
Summary
FDA records medical device Class II recall Z-2470-2012, published October 3, 2012 and initiated May 29, 2012. Recalling firm: Alcon Research, LTD.; Status: Terminated; Reported quantity: 93,319 (30, 413 distributed domestically & 62,906 internationally); Lot: 1) 23G TotalPLUS Pak: Accurus Standalone - 8065750828, 2) 23G Combined Pack 0.9mm: Accurus Standalone - 8065750814, 3) 23G Combined Pack 1.1mm: Accurus Standalone - 8065750816, 4)23G TotalPLUS Pak: Constellation Standalone - 8065751058, 5) 23G Combined Pack 0.9mm: Constellation Standalone -8065751077, 6) 23G Combined Pack 1.1mm: Constellation Standalone - 8065751065. 7) 23G Small Parts Kit - 8065750842. Complete list of lot numbers will be provided with recall recommendation. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
This medical device correction is related to Alcon products that present a 23G non-valved trocar cannula and infusion cannula together because of an issue regarding the loose fit between the 23G non-valved trocar cannula and infusion cannula that could potentially cause a detachment between the non-valved trocar cannula and infusion cannula during surgery.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Alcon Research, LTD.
- Distributor
- Worldwide Distribution - USA (nationwide) and the countries of Saudi Arabia, United Kingdom, Greece, Singapore, Germany, Switzerland, Kuwait, United Arabian Emirates, and Austria.
- Lot
- 1) 23G TotalPLUS Pak: Accurus Standalone - 8065750828, 2) 23G Combined Pack 0.9mm: Accurus Standalone - 8065750814, 3) 23G Combined Pack 1.1mm: Accurus Standalone - 8065750816, 4)23G TotalPLUS Pak: Constellation Standalone - 8065751058, 5) 23G Combined Pack 0.9mm: Constellation Standalone -8065751077, 6) 23G Combined Pack 1.1mm: Constellation Standalone - 8065751065. 7) 23G Small Parts Kit - 8065750842. Complete list of lot numbers will be provided with recall recommendation.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDALegion FMS PakAugust 23, 2023
- FDACyPass System 241-S, CYPASS MICRO-STENT, CYPASS LOADED, CYPASS APPLIER, REF CATALOG #: 8065754003 ForOctober 31, 2018
- FDAAlcon Custom PakNovember 8, 2017
- FDACatheter Irrigation Set labeled in part: ***24K Arthroscopy Outflow/Suction Tubing Set*** This ArthroscopyOctober 3, 2012
- FDAUltraFlow HPC Flow Directed Micro Catheter, Model # 105-5065. Product Usage: The UltraFlow HPC Flow DirectedOctober 3, 2012
- FDATSX-301A/2 Aquilion One CT System. The TSX-30IAl2 is a whole body CT scanner. This device captures crossOctober 3, 2012
Questions about this recall
Was This voluntary medical device correction is related to Alcon products that present a 23G nonvalved trocar cannula and infusion cannula together. These items are available as components within Accurus 23G TotalPLUS Pak, Accurus 23G Combined Procedure Packs (0.9mm and 1.1mm), Constellation 23G Total PLUS Pak, Constellation 23G Combined Pack (0.9 mm and 1.1 mm) and 23G Small Parts Kit. Single use 23G non-valved trocar cannulas are used to provide internal access to the ocular cavity during ophthalmic surgery. Single use infusion cannulas are used for the infusion of fluid or air during ophthalmic surgery to maintain intraocular pressure during posterior segment surgeries. recalled?
Yes. FDA published recall Z-2470-2012 on October 3, 2012. Reported hazard: This medical device correction is related to Alcon products that present a 23G non-valved trocar cannula and infusion cannula together because of an issue regarding the loose fit between the 23G non-valved trocar cannula and infusion cannula that could potentially cause a detachment between the non-valved trocar cannula and infusion cannula during surgery.
What should I do if I have this product?
Voluntary: Firm initiated · Letter