00780302801 MIS 12/14 Slotted Femoral Head Provisional -3.5X28 00780302802 MIS 12/14 Slotted Femoral Head Provisional +0X28 00780302803 MIS 12/14 Slotted Femoral Head Provisional +3.5X28 00780302804 MIS 12/14 Slotted Femoral Head Provisional +7X28 00780302805 MIS 12/14 Slotted Femoral Head Provisional +10.5X28 00780302814 MIS 12/14 Slotted Femoral Head Provisional +7X28 The provisional heads replicate final femoral implant geometry and are used to trial range of motion, joint stability, and determine appropriate femoral head size and offset during total hip arthroplasty.
Published October 10, 2012 · Recall initiated: April 1, 2011
Summary
FDA records medical device Class II recall Z-2481-2012, published October 10, 2012 and initiated April 1, 2011. Recalling firm: Zimmer, Inc.; Status: Terminated; Reported quantity: 231,641 all lots, all sizes; Lot: All lots including lot 78939000 (1/7/2002) to present. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Complaints have been received reporting sub-optimal fit between provisional femoral heads (VerSys and MIS) and the mating components (ie stems and provisional femoral necks), in some cases alleging the fit was too loose and in other cases, alleging the fit was too tight. Intraoperative dislodging of the provisional head and the provisional head being left in-vivo after separating from mating comp
What to do
Voluntary: Firm initiated · E-Mail
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Zimmer, Inc.
- Distributor
- Worldwide Distribution - USA (nationwide) and internationally to Canada, Mexico, South America, Europe, Puerto Rico, China, and Korea.
- Lot
- All lots including lot 78939000 (1/7/2002) to present
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAOff-Axis Comprehensive Shoulder System Small, Augment, Off-Axis, Reamer Guide, Model/Catalog Number: 110040240July 1, 2026
- FDABrand Name: Persona Revision Product Name: Persona Revision® Trabecular Metal® Femoral Distal AugmentJuly 1, 2026
- FDAOff-Axis Alliance Glenoid, Left, 4-Peg, Augment, Reamer Guide, Model/Catalog Number: 110040444July 1, 2026
- FDAOff-Axis Comprehensive Shoulder System, Medium, Augment, Off-Axis, Reamer Guide Model/Catalog Number:July 1, 2026
- FDA00780302601 MIS 12/14 Slotted Femoral Head Provisional -3.5X26 00780302602 MIS 12/14 Slotted Femoral HeadOctober 10, 2012
- FDA00780303201 MIS 12/14 Slotted Femoral Head Provisional -3.5X32 00780303202 MIS 12/14 Slotted Femoral HeadOctober 10, 2012
Questions about this recall
Was 00780302801 MIS 12/14 Slotted Femoral Head Provisional -3.5X28 00780302802 MIS 12/14 Slotted Femoral Head Provisional +0X28 00780302803 MIS 12/14 Slotted Femoral Head Provisional +3.5X28 00780302804 MIS 12/14 Slotted Femoral Head Provisional +7X28 00780302805 MIS 12/14 Slotted Femoral Head Provisional +10.5X28 00780302814 MIS 12/14 Slotted Femoral Head Provisional +7X28 The provisional heads replicate final femoral implant geometry and are used to trial range of motion, joint stability, and determine appropriate femoral head size and offset during total hip arthroplasty. recalled?
Yes. FDA published recall Z-2481-2012 on October 10, 2012. Reported hazard: Complaints have been received reporting sub-optimal fit between provisional femoral heads (VerSys and MIS) and the mating components (ie stems and provisional femoral necks), in some cases alleging the fit was too loose and in other cases, alleging the fit was too tight. Intraoperative dislodging of the provisional head and the provisional head being left in-vivo after separating from mating comp
What should I do if I have this product?
Voluntary: Firm initiated · E-Mail