MEDLINE Medical Procedure Kits labeled as: 1) KIT UROLOGY URETHROPLASTY, Medline Kit Number/SKU DYKMBNDL117C; 2) URETHROPLASTY, Medline Kit Number/SKU DYNJ908978; 3) URETHROPLASTY, Medline Kit Number/SKU DYNJ908978F.
Published June 24, 2026 · Recall initiated: May 4, 2026
Summary
FDA records medical device Class II recall Z-2495-2026, published June 24, 2026 and initiated May 4, 2026. Recalling firm: Medline Industries, LP; Status: Ongoing; Reported quantity: 91 units; Lot: Medline Kit Number/SKU DYKMBNDL117C: UDI/DI 10195327371104 (each), 40195327371105 (case), Lot Number 23DLA834; Medline Kit Number/SKU DYNJ908978: UDI/DI 10195327231132 (each), 40195327231133 (case), Lot Number 23BMC463; Medline Kit Number/SKU DYNJ908978: UDI/DI 10195327231132 (each), 40195327231133 (case), Lot Number 23AMA685; Medline Kit Number/SKU DYNJ908978: UDI/DI 10195327231132 (each), 40195327231133 (case), Lot Number 23AMB905; Medline Kit Number/SKU DYNJ908978: UDI/DI 10195327231132 (each), 40195327231133 (case), Lot Number 22LMG466; Medline Kit Number/SKU DYNJ908978: UDI/DI 10195327231132 (each), 40195327231133 (case), Lot Number 22LMB983; Medline Kit Number/SKU DYNJ908978F: UDI/DI 10198459228803 (each), 40198459228804 (case), Lot Number 25JBD225; Medline Kit Number/SKU DYNJ908978F: UDI/DI 10198459228803 (each), 40198459228804 (case), Lot Number 25HBW236. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
What to do
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Medline Industries, LP
- Distributor
- Worldwide distribution - US Nationwide and the countries of AE, BB, BM, BS, CA, CY, GU, NL, PA, QA, SA, VI.
- Lot
- Medline Kit Number/SKU DYKMBNDL117C: UDI/DI 10195327371104 (each), 40195327371105 (case), Lot Number 23DLA834; Medline Kit Number/SKU DYNJ908978: UDI/DI 10195327231132 (each), 40195327231133 (case), Lot Number 23BMC463; Medline Kit Number/SKU DYNJ908978: UDI/DI 10195327231132 (each), 40195327231133 (case), Lot Number 23AMA685; Medline Kit Number/SKU DYNJ908978: UDI/DI 10195327231132 (each), 40195327231133 (case), Lot Number 23AMB905; Medline Kit Number/SKU DYNJ908978: UDI/DI 10195327231132 (each), 40195327231133 (case), Lot Number 22LMG466; Medline Kit Number/SKU DYNJ908978: UDI/DI 10195327231132 (each), 40195327231133 (case), Lot Number 22LMB983; Medline Kit Number/SKU DYNJ908978F: UDI/DI 10198459228803 (each), 40198459228804 (case), Lot Number 25JBD225; Medline Kit Number/SKU DYNJ908978F: UDI/DI 10198459228803 (each), 40198459228804 (case), Lot Number 25HBW236.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAMEDLINE medical convenience kits labeled as: 1) HIP SCOPE PACK, Medline Kit Number/SKU DYNJ64103C; 2) HIPOctober 28, 2026
- FDAMEDLINE medical convenience kits labeled as: 1) BIRTHDAY KIT-DOUBLE BASIN-LF, Medline Kit Number/SKUOctober 28, 2026
- FDAMEDLINE DRAPE C-SECTION FEN CLR SCREEN W, Medline Product Number/SKU DYNJP6108October 28, 2026
- FDAMEDLINE medical convenience kits labeled as: FETOSCOPY CDS-LF, Medline Product Number/SKU CDS840215POctober 28, 2026
- FDAMEDLINE Medical Procedure Kits labeled as: TONSIL PACK CATH, Medline Kit Number/SKU DYNJ65199AJune 24, 2026
- FDAMEDLINE Medical Procedure Kits labeled as: 1) CV DRAPING/BASIN, Medline Kit Number/SKU DYNJ908483A; 2) CVJune 24, 2026
Questions about this recall
Was MEDLINE Medical Procedure Kits labeled as: 1) KIT UROLOGY URETHROPLASTY, Medline Kit Number/SKU DYKMBNDL117C; 2) URETHROPLASTY, Medline Kit Number/SKU DYNJ908978; 3) URETHROPLASTY, Medline Kit Number/SKU DYNJ908978F. recalled?
Yes. FDA published recall Z-2495-2026 on June 24, 2026. Reported hazard: During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
What should I do if I have this product?
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit