Brand Name: Inspire Medical Systems, Inc. Product Name: Inspire stimulation Lead Model/Catalog Number: 4063 Product Description: The Inspire® Stimulation Lead (Model 4063) is designed to deliver stimulation to the hypoglossal nerve for the treatment of obstructive sleep apnea. The lead features a flexible, self-sizing cuff. Electrodes in the inner surface of the cuff deliver stimulation to the nerve. The lead incorporates a standard connector for coupling to the implantable pulse generator (IPG). Component: N/A
Published June 24, 2026 · Recall initiated: May 11, 2026
Summary
FDA records medical device Class II recall Z-2516-2026, published June 24, 2026 and initiated May 11, 2026. Recalling firm: Inspire Medical Systems Inc.; Status: Ongoing; Reported quantity: 10 impacted units; Lot: Lot Code: Model 4063 (USA only lead); UDI-DI [4063] 10855728005793; Inspire part number [4063] 900-013-001; serial numbers D53393, D61104 Model 4063 (Global lead); UDI-DI [4063] 10810098650295, Inspire part number [4063] 900-013-003; serial numbers D89539, D89540, D89543, D89559, D89563, D89564, D89577, D8959. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Products are labeled with incorrect Use By Date, extending beyond the validated three-year shelf life from date of manufacturer. Material degradation of a product past its Use By Date may result in unplanned or premature revision surgery.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Inspire Medical Systems Inc.
- Distributor
- Worldwide distribution - US Nationwide and the country of Hong Kong.
- Lot
- Lot Code: Model 4063 (USA only lead); UDI-DI [4063] 10855728005793; Inspire part number [4063] 900-013-001; serial numbers D53393, D61104 Model 4063 (Global lead); UDI-DI [4063] 10810098650295, Inspire part number [4063] 900-013-003; serial numbers D89539, D89540, D89543, D89559, D89563, D89564, D89577, D8959
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDABrand Name: Inspire Medical Systems, Inc. Product Name: Inspire Respiratory Sensing Lead Model/Catalog Number:June 24, 2026
- FDASplashwire Hydrophilic Guide Wire: REF: MSWSTDA35150 UDI-DI codes: 00884450529645 00884450529669June 24, 2026
- FDABD Connecta" BD Luer-Lok" 360 REF 394910 UDI-DI code: 00382903949106 BD Connecta" Stopcocks are used toJune 24, 2026
- FDAInspire Model 4340 Respiratory Sensing Lead, component of the Inspire Upper Airway Stimulation (UAS) system.October 16, 2024
- FDAInspire Model 4063 Stimulation Lead, component of the Inspire Upper Airway Stimulation (UAS) system.October 16, 2024
- FDAInspire Model 3028, IV Implantable Pulse GeneratorJuly 3, 2024
Questions about this recall
Was Brand Name: Inspire Medical Systems, Inc. Product Name: Inspire stimulation Lead Model/Catalog Number: 4063 Product Description: The Inspire® Stimulation Lead (Model 4063) is designed to deliver stimulation to the hypoglossal nerve for the treatment of obstructive sleep apnea. The lead features a flexible, self-sizing cuff. Electrodes in the inner surface of the cuff deliver stimulation to the nerve. The lead incorporates a standard connector for coupling to the implantable pulse generator (IPG). Component: N/A recalled?
Yes. FDA published recall Z-2516-2026 on June 24, 2026. Reported hazard: Products are labeled with incorrect Use By Date, extending beyond the validated three-year shelf life from date of manufacturer. Material degradation of a product past its Use By Date may result in unplanned or premature revision surgery.
What should I do if I have this product?
Voluntary: Firm initiated · Letter