Philips IntelliSpace ECG Management System, with software option C61 needed to import ECGs from the Quinton Stress ECG System. Model No. 860426
Published August 27, 2014 · Recall initiated: August 7, 2014
Summary
FDA records medical device Class II recall Z-2535-2014, published August 27, 2014 and initiated August 7, 2014. Recalling firm: Philips Medical Systems, Inc.; Status: Terminated; Reported quantity: 11 units; Lot: Model No. 860426 Serial Numbers: US11400027, US31400041, US31400042, US31400044, US31400045, US31400046, US41400051, US51300008, US51400056, US81300016, USN1300020. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Philips IntelliSpace ECG (IECG) Management Systems that are importing Stress ECG records from a Quinton Stress ECG system have the potential for the patient record to contain multiple and therefore inaccurate patient identifiers such as: patient name, patient ID (PID), Medical record number (MRN). This creates the potential for misdiagnosis and incorrect therapy.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Philips Medical Systems, Inc.
- Distributor
- Worldwide Distribution - US including the states of CA, GA, NC, NH, NY and OR., and the countries of Canada and Spain.
- Lot
- Model No. 860426 Serial Numbers: US11400027, US31400041, US31400042, US31400044, US31400045, US31400046, US41400051, US51300008, US51400056, US81300016, USN1300020
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAFoot Switches used with Philips MultiDiagnost Eleva with Flat Detector; Product Codes: 708037, 708038 ThisApril 6, 2016
- FDAPhilips Healthcare Xper vascular systems R7.6 Model : 722134 Product Usage: The Allura Xper series is intendedDecember 9, 2015
- FDAPhilips Healthcare INTEGRIS vascular System Model : 722122 Product Usage: The Allura Xper series is intendedDecember 9, 2015
- FDAPhilips Healthcare INTEGRIS H5000F/Allura 9F Model : 722017 Product Usage: The Allura Xper series is intendedDecember 9, 2015
- FDAFamotidine for Oral Suspension, USP 40 mg/5 mL, Rx Only, Cherry-Banana-Mint Flavored, Not for Injection,September 3, 2014
- FDAPersona (TM) The Personalized Knee System CPS Tibial Articular Surface Provisional Top Nonsterile TASP MLC LAugust 27, 2014
Questions about this recall
Was Philips IntelliSpace ECG Management System, with software option C61 needed to import ECGs from the Quinton Stress ECG System. Model No. 860426 recalled?
Yes. FDA published recall Z-2535-2014 on August 27, 2014. Reported hazard: Philips IntelliSpace ECG (IECG) Management Systems that are importing Stress ECG records from a Quinton Stress ECG system have the potential for the patient record to contain multiple and therefore inaccurate patient identifiers such as: patient name, patient ID (PID), Medical record number (MRN). This creates the potential for misdiagnosis and incorrect therapy.
What should I do if I have this product?
Voluntary: Firm initiated · Letter