H4051 Endo Tube Holder, Pedi 2.5mm Pediatric Endotracheal Intubation Kit

Published September 3, 2014 · Recall initiated: July 7, 2014

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The tube fastener on the ET Tube Holder may compress the tracheal tube when used which may cause difficulty when locking the ET Tube Holder in place or difficulty when passing a suction catheter.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Smiths Medical ASD, Inc.
Distributor
Worldwide Distribution - USA including AK, AR, AZ, CA, CO, CT, DC, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV and Internationally to CA, CY, GB, IE, IT, JE, KW, MT, PL, RU, SA, ZA.
Lot
Lot Numbers: 2597332, 2585074, 2580798, 2576365, 2564404, 2564403, 2563088, 2549367, 2519413, 2500170, 2488207, 2474646, 2465828, 2453473, 2443378, 2438615, 2433450, 2412145, 2380351, 2376498, 2365494, 2365493, 2361580, 2357823, 2353347, 2343313, 2329583, 2326899, 2322825, 2298752, 2286594, 2278069, 2271358, 2263553, 2260562, 2260561, 2250053, 2250052, 2250050, 2250049, 2244749, 2228363, 2219213, 2211929, 2208194, 2191930, 2179229, 2175056, 2154534, 2142178, 2135983, 2128635, 2124241, 2120119, 2116099, 2090223, 2077179, 2063989, 2063988, 12040, 12033, 11877, 11876, 11874, 11873, 11679, 11655, 11407, 11405
Risk classification
Class I
Area (FDA)
device
Country
United States