Knee Instrument Fixation Screws: 48MM HEADED SCREW 48MM HEADLESS SCREW MIS HEADED SCREW, 48MM MIS HEADLESS SCREW, 48MM

Published September 3, 2014 · Recall initiated: July 22, 2014

Summary

FDA records medical device Class II recall Z-2546-2014, published September 3, 2014 and initiated July 22, 2014. Recalling firm: Zimmer, Inc.; Status: Terminated; Reported quantity: 972,125 units; Lot: Item Numbers 00579104100, 00579104200, 00598304048, 00598304148 Exp Date Prior to 2023-11. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Zimmer is recalling sterile-packaged implants and instruments packaged in six different package configurations due to packaging design verification test failures. Specifically, multiple test samples from each of the six configurations failed simulated distribution and shipping testing when the devices either breached inner or outer package trays or seal integrity was lost.

What to do

Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Zimmer, Inc.
Distributor
Worldwide Distribution.
Lot
Item Numbers 00579104100, 00579104200, 00598304048, 00598304148 Exp Date Prior to 2023-11
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Knee Instrument Fixation Screws: 48MM HEADED SCREW 48MM HEADLESS SCREW MIS HEADED SCREW, 48MM MIS HEADLESS SCREW, 48MM recalled?

Yes. FDA published recall Z-2546-2014 on September 3, 2014. Reported hazard: Zimmer is recalling sterile-packaged implants and instruments packaged in six different package configurations due to packaging design verification test failures. Specifically, multiple test samples from each of the six configurations failed simulated distribution and shipping testing when the devices either breached inner or outer package trays or seal integrity was lost.

What should I do if I have this product?

Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit