Summary

FDA records medical device Class II recall Z-2558-2025, published September 24, 2025 and initiated July 8, 2025. Recalling firm: Boston Scientific Neuromodulation Corporation; Status: Ongoing; Reported quantity: 25,260; Lot: UDI-DI: 08714729985044, 08714729985051. Document Number/Revision: 92328632-09, Rev A. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Wire break(s) have occurred in rechargeable deep brain stimulation implantable pulse generators, implanted in the submuscular pectoral location. Wire breaks prevent successful stimulation, thus requiring device removal/replacement and users may experience high monopolar impedances, undesired sensation, loss of therapy, return of symptoms and/or Bluetooth connectivity challenges.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Boston Scientific Neuromodulation Corporation
Distributor
Worldwide - US Nationwide distribution including in the states of MN, FL, CO, MD, NY, NE, PA, WI, AZ, TN, MO, TX, MI, MT, MA, WV, AL, VA, NC, CA, OR, WA, LA, OH, NJ, NH, SD, GA, CT, NV, UT, IN, ID, IL, ME, SC, DC, OK, KY, AK, HI, RI, AR, KS, NM and the countries of Argentina, Brazil, Chile, Ecuador, Colombia , Mexico, Canada, Japan, Algeria, Austria, Azerbaijan, Belgium, Bulgaria, Croatia, Denmark, Egypt, Finland, France, Georgia, Germany, Greece, Hungary, Iceland, Iran, Ireland, Israel, Italy, Jordan, Kazakhstan, Kuwait, Lebanon, Libya, Luxembourg, Malta, Morocco, Netherlands, Norway, Pakistan, Poland, Portugal, Romania, Russian Fed., Saudi Arabia, Serbia, Slovakia, South Africa, Spain, Sweden, Switzerland, Turkiye, UAE, Ukraine, United Kingdom, Hong Kong, China, Singapore, Korea, Taiwan , India, Thailand.
Lot
UDI-DI: 08714729985044, 08714729985051. Document Number/Revision: 92328632-09, Rev A
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Vercise Deep Brain Stimulation Systems Surgical Implant Manual, Document Number: 92328632-09 recalled?

Yes. FDA published recall Z-2558-2025 on September 24, 2025. Reported hazard: Wire break(s) have occurred in rechargeable deep brain stimulation implantable pulse generators, implanted in the submuscular pectoral location. Wire breaks prevent successful stimulation, thus requiring device removal/replacement and users may experience high monopolar impedances, undesired sensation, loss of therapy, return of symptoms and/or Bluetooth connectivity challenges.

What should I do if I have this product?

Voluntary: Firm initiated · Letter