Octopus 4 Tissue Stabilizer, Model 29400

Published July 1, 2026 · Recall initiated: May 12, 2026

Summary

FDA records medical device Class II recall Z-2577-2026, published July 1, 2026 and initiated May 12, 2026. Recalling firm: Medtronic Perfusion Systems; Status: Ongoing; Reported quantity: 590 units; Lot: UDI-DI: 00763000543679; Serial Numbers: 0231266463 0231468923 0231478249 0231478253 0231478260 0231651627 0231651634 0231651636 0231651637 0231651638 0231651639 0231651641 0231651642. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

During the manufacturing process, Medtronic personnel identified an assembly issue where the tubing was incorrectly positioned. The short tube was attached to the bottom of the canister, and the long tube was attached to the top, contrary to the intended design.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Medtronic Perfusion Systems
Distributor
Worldwide - US Nationwide distribution.
Lot
UDI-DI: 00763000543679; Serial Numbers: 0231266463 0231468923 0231478249 0231478253 0231478260 0231651627 0231651634 0231651636 0231651637 0231651638 0231651639 0231651641 0231651642
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Octopus 4 Tissue Stabilizer, Model 29400 recalled?

Yes. FDA published recall Z-2577-2026 on July 1, 2026. Reported hazard: During the manufacturing process, Medtronic personnel identified an assembly issue where the tubing was incorrectly positioned. The short tube was attached to the bottom of the canister, and the long tube was attached to the top, contrary to the intended design.

What should I do if I have this product?

Voluntary: Firm initiated · Letter